8
MAX
30V
30V
MAX
MAX
30V
INPUT
OUTPUT
COM
mA
V
+LOOP
XMT
+mA-
+
Figure 5.1
Current output
5.3.2
25% step current output
•
Connect the test leads to the device as in Figure 5.1.
•
Press the output/input toggle button (3) so that the OUTPUT symbol appears on the
display. The device is now in an output state.
•
Press the 25%/RAMP button, the
or
▲ symbols will be displayed.
•
Press the mA%/V unit select button (4) so the device displays mA or mA%.
•
Use the
▲ or ▼ buttons to change the output value in 25% increments; where 0% is
4mA and 100% is 20mA.
•
Press the 25%/RAMP button again to exit the step current output.
5.3.3
Current Output Set for Zero Point & Full Scale
•
Connect the test leads to the device as in Figure 5.1.
•
Press the output/input toggle button (3) so that the OUTPUT symbol appears on the
display. The device is now in an output state.
•
Press the 100%/START button and the
,
▲, 0 and FS symbols will be displayed.
•
Use the
▲ or ▼ buttons to change the output value from 100% (20mA) and 0% (4mA).
•
Press the 100%/START button again to exit the step current output.