5.3.4 

Auto ramp output 

 

Connect the test leads to the device as in Figure 5.1. 

 

Press the output/input toggle button (3) so that the  OUTPUT symbol appears on the 
display. The device is now in an output state. 

 

Press the STEP/AUTO button, the OUTPUT, OFF and 4mA symbols will be displayed. 
The device is now in ramp mode. 

 

Press the 25%/RAMP key again to change the type of output slope, which is displayed 
at the lower left of the display. The types are displayed as 

Ʌ S

Ʌ F

 in the correct 

order. These symbols indicate low-speed ramps and high-speed ramps, respectively. 
The former has a maximum cycle of 60s, the latter has a maximum cycle of 30s, and 
the auto-step ramp pauses for 5s at each step. 

 

When the  ON symbol is displayed, press the 100%/START button to start setting the 
waveform output. 

 

Press the  100%/START  button  again,  the  output  will  be  paused  at  the current value 
and the OFF symbol will be displayed. 

 

Press the button again, to continue the stepped output from the paused value. 

 

When the  OFF symbol appears, press any of the buttons 

▲, ▼, ◄, ► to return the 

output to 0%. Then, the value of 4 mA will be displayed. 

5.3.5 

Simulating transmitter output (XMT) 

 

Insert one end of the test lead to the XMT output jack of the device and connect the 
other end with the input of the device as shown in Figure 5.2. 

 

The key-operation is the same as that of the current output. 

 

 

MAX

30V

30V

 

MAX

 

MAX

30V

INPUT

OUTPUT

COM

mA

 V

+LOOP

XMT

+mA-

+

5--25VDC Power

 

Power

 

Figure 5.2 

XMT output connection 

 

 

 

Note 

 

Range of power supply: 5 to 25VDC 

 

Connect a 24V DC power supply to the device while  operating the current output, to 
prolong the working life of the battery. 

 

 

Содержание PCE-LOC 10

Страница 1: ...oop Calibrator User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last change...

Страница 2: ...3 3 2 LCD layout 4 4 Device Maintenance 4 4 1 General maintenance 4 4 2 Battery replacement 5 4 3 Fuse replacement 6 5 Operating Instructions 6 5 1 Power on off 6 5 2 Automatic power off 7 5 3 Device...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...4V DC 25 mA 10 Voltage measurement Area Resolution Accuracy 0 2 28V 1 mV 0 02 of meas 2 mV Current measurement Area Resolution Accuracy 1 22 mA 0 001 mA 0 02 of meas 4 A Current measurement Loop curre...

Страница 5: ...evice layout 1 LCD Display 6 25 single step Auto ramp 2 Power 7 0 100 of setting Auto ramp start 3 Output Input toggle 8 Output value set 4 mA unit select 9 Input digit select 5 Single step Auto toggl...

Страница 6: ...input and output values ON OFF These symbols denote the status of the input output signals These symbols denote the high and low speed of the auto ramp and auto step ramp 4 Device Maintenance This se...

Страница 7: ...olarity of the batteries with the polarity on the device Remove the batteries if the device is not to be used for a prolonged period of time Dispose of the old batteries according to local regulations...

Страница 8: ...with a screwdriver Replace the blown fuse s Reinstall the cover Reinstall the device protector Figure 3 2 Fuse replacement 5 Operating Instructions 5 1 Power on off Press the power button to turn the...

Страница 9: ...ution Do not apply voltage to the output terminal during the operation If any improper voltage is applied to the output terminal it will cause damage to the internal circuitry Function Operation Opera...

Страница 10: ...vice displays mA or mA Use the or buttons to change the output value in 25 increments where 0 is 4mA and 100 is 20mA Press the 25 RAMP button again to exit the step current output 5 3 3 Current Output...

Страница 11: ...ol is displayed press the 100 START button to start setting the waveform output Press the 100 START button again the output will be paused at the current value and the OFF symbol will be displayed Pre...

Страница 12: ...V 28 V 20 mA 00 000 mA 25 00 mA 0 000 V 1 000 22 000 mA 31 25 112 50 mA 0 2000 28 000 V 5 4 1 DC current measurement Insert one end of the test lead to the mA jack INPUT of the device and connect the...

Страница 13: ...symbol appears on the display The device is now in an input state Press the mA V unit select button 4 so the device displays V The device now begins to measure ON and the measured result are now show...

Страница 14: ...ge Output Range Resolution Accuracy Remark DCA 20mA 0 000 22 000mA 0 001mA 0 05 set value 4uA Max load 1K at 20mA Simu transmitter absorption current 20mA 0 000 22 000mA 0 001mA 0 05 set value 4uA Max...

Страница 15: ...isposal of batteries in the EU the 2006 66 EC directive of the European Parliament applies Due to the contained pollutants batteries must not be disposed of as household waste They must be given to co...

Страница 16: ...583 975 114 Fax 49 0 2903 976 99 9971 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce instruments com info pce instruments co uk info pce italia it www pce instruments com deutsch www pce instrumen...

Отзывы: