3-2 Electrical Specifications
Unless otherwise stated, the following specifications hold under
the following conditions :
•
The meter is located in the far-field of a source, the sensor head is
pointed towards the source.
•
Ambient temperature : +23
℃
±
3
℃
•
Relative air humidity : 25% to 75%
Sensor type :
Electrical field (E)
Frequency range :
10MHz to 8GHz
Specified measurement range :
•
CW signal (f > 50MHz)
: 20mV/m to 108.0V/m ,
53
µ
A/m to 286.4mA/m ,
1
µ
W/m
2
to 30.93W/m
2
,
0
µ
W/cm
2
to 3.093mW/cm
2
Dynamic range :
Typically 75dB
Absolute error at 1 V/m and 50 MHz :
±
1.0dB
Frequency response :
•
Sensor taking into account the typical CAL factor
:
±
1.0dB (50MHz to 1.9GHz)
±
2.4dB (1.9GHz to 8GHz)
•
Isotropy deviation
: Typically
±
1.0dB (f>50MHz)
•
Overload limit
: 10.61mW/cm
2
(200V/m)
•
Thermal response (0 to 50
℃
)
:
±
0.5dB