Manual
10
If any samples are assigned to a standard, you can view them by pressing Enter
.
If there is more than one sample, you can use the arrow keys (
and
) for selection.
If you want to delete a sample, select it and hold the Back button
for approx. 3 seconds.
If you press Back
briefly, you get back to the main screen of the record function.
If you want to add samples to an existing standard, select the standard and press the Menu
button
. You will now be redirected to the standard measurement screen. Press Enter
to go
to the sample measurement screen. Now you can perform a sample measurement which will
automatically be added to the selected standard.
If you want to turn a sample into a standard, select the standard which the sample is assigned to
and press Enter
. Now, select the sample and press the Menu button
. The sample has now
turned into a standard.
5.3.2 Calibration
Auto calibration
During start-up, the PCE-CSM 7 automatically performs a white and black calibration. For this purpose,
the device comes with a calibration cover which should be used at every startup. You just have to place
the device on top of it and make sure that the calibration plate is located under the measuring aperture.
If the auto calibration was successful, you automatically get to the main screen (standard measurement).
You can now remove the calibration cover.
If the auto calibration was not successful, the display shows a screen where you have the option to restart
the white calibration or to skip it. If you choose skip, the PCE-CSM 7 uses the last white and black
calibration.
We do not recommend skipping the auto calibration.
Manual calibration
To get to the calibration menu, press the Menu button
, select “Calibrate“ and press Enter .
Here you can select between white and black calibration. Use the arrow keys (
and
) to make a
selection and press Enter
to confirm. A confirmation screen with instructions will appear.
White calibration
If you want to perform a white calibration, place the device on the calibration cover. Make sure that the
sensor is located above the calibration plate. After that, press the Testing button to start the calibration.
Black calibration
If you wish to perform a black calibration, remove the calibration plate and hold the device with the
measuring aperture pointing upwards. Make sure that this calibration is carried out in a dark environment.