Manual
10
c) Setup Details
(1) Press S/M to select the desired option.
(2) The desired value can be entered when an appropriate option is in a reverse black / white mode.
(3) The Mapping unit for 4mA and 20mA must be entered with the help of the ASCII code. (See table
Annex 1), e.g., the codes "75"
and “87” must be entered for "KW".
(4) MAPPING:
“YES” mapping function activated
“NO” mapping function deactivated
d) MAPPING Function
(1) The mapping function is activated when the user has selected "YES" in MAPPING in the SETUP
Menu.
(2) The display unit then corresponds to the unit set by the user in the SETUP menu (if "0KW" for
4mA and "100KW" for 20mA were set in the SETUP menu).
(3) If 100 has been typed by the user and ENTER pressed, the display shows the following values:
100.00KW (on the main display) and 20.000mA (the current value output by the PCE-789).
(4) If the ramp function is used, the display shows 0 ~ 100KW instead of 4 ~ 20mA.
5.3
Temperature Source
a) Simulation of the Output Signals of the Thermocouples
(for the types K, J, E, T, R, S, N, L, U, B, C, and mV output)
(1)
Turn on the device and push the slide switch to the “°C ° F mV”.
(2) Select a thermocouple type in the Setup menu.
(3) Enter a temperature value (incl. digits after the decimal point) and then press ENTER.
(4) Connect the thermocouple cable to the TC/mV jack.
(5) Finally, connect the thermocouple cable to the calibrated object.