Operation 

5.1 

Measurement 

Place  the  device  on the sample  and  push  it  down  on  the surface  with  enough  pressure.  Read 
out the hardness value from the display. 

5.2 

Further functions 

5.2.1 

Max hold function 

Press  the  Maximum  /  Plus  key.  The 

‘MAX’  indication  appears  in  the  upper  right corner  of the 

display. The maximum value since the activation of the function is shown on the display. To exit 
the max hold function, just press the Power / Menu Key. T

he ‘MAX’ indication disappears. 

5.2.2 

Average function 

Press  the  Average  /  Minus  key.  The  statistical  measurement  number  indication 

‘N’  and  the 

‘MAX’ indication appear on the display. Additionally, a ‘0’ appears on the left side of the display 
and 

‘0.0’ appears on the right side. The left value is the number of measurements taken. The 

right value is the reading of the last measurement.  
Take  measurements  until  the  number  of  measurements  taken  reaches  the  set  value  of  the 
statistical  measurement  numb

er  ‘N’.  Now  the  device  calculates  the  average  value  of  the 

measurements taken. The result is shown on the right side of the display. In addition, an ‘AVE’ 
indication appears and the device emits an acoustic signal. 

5.2.3 

Setting the statistical measurement n

umber ‘N’ 

To set  the statistical measurement  number,  press  and  hold  the Power  /  Menu  key  for  about  6 

seconds  until 

the  ‘

 

’  indication  appears  on the  display. Press the Maximum  /  Plus key  or 

Average / Minus key to adjust the measurement number. The statistical measurement number 
can  be  set  between  1~29.  Press  the  Power  /  Menu  Key  to  return  the  standard  measurement 
mode. 
 
Use  the  method  of  calculating  the  average  of  several  measurements  to  figure  out  the 
measurement value. The softer the sample is, the more measurements should be included. The 
suggested number of measurements can be seen in the following tables: 
 
Suggested number of measurements for aluminium alloys  
(According to ASTM B648-2000) 
 
Barcol hardness value 

Min. number of measurements 

50 

60 

70 

80 

 
 

 

Содержание PCE-1000N

Страница 1: ...N Barcol Hardness Tester User Manual User manuals in various languages Deutsch fran ais italiano espa ol portugu s nederlands t rk polski can be downloaded here www pce instruments com Last change 13...

Страница 2: ...4 3 3 Display 5 4 Getting started 5 4 1 Power supply 5 4 2 Preparation 5 5 Operation 6 5 1 Measurement 6 5 2 Further Functions 6 6 Calibration 7 6 1 Low end calibration 7 6 2 High end calibration 7 6...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...ution 0 1 HBa Measurement error 81 88 HBa 1 HBa 42 48 Hba 2 HBa Repeatability error 81 88 HBa 1 5 HBa 42 48 HBa 2 5 Hba Operating conditions temperature humidity 0 50 C 50 RH Power supply 2 x 1 5 V AA...

Страница 5: ...ead The effective measuring range is equivalent to a Brinell hardness of 25 150HB The max hold function shows the maximum value during the measurement The PCE 1000N can be used for measuring not only...

Страница 6: ...0076 HBa Barcol hardness value H indentation depth mm 0 0076 indentation depth at 1 Barcol 2 1 housing screws 2 2 high end calibration button 2 3 low end calibration button 2 4 spring tube 2 5 lower s...

Страница 7: ...lat surface and take a measurement The reading should be in the indicated range of two hardness blocks If it is out of the range a calibration according to chapter 6 is needed The sample surface shoul...

Страница 8: ...asurements taken reaches the set value of the statistical measurement number N Now the device calculates the average value of the measurements taken The result is shown on the right side of the displa...

Страница 9: ...se a small screw driver to loosen the main spindle adjusting screw in counter clockwise direction until the end Put the pin tube into the hole of the pin length calibration block and push the tester s...

Страница 10: ...th a small screwdriver a total of 12 Take off the case and take out the spring tube Loosen the lower spindle connecting screw with a spanner and take out the lower spindle Now you can remove the used...

Страница 11: ...version between different hardness systems is not possible Therefore the following conversion table is for reference only It is recommended to determine the actual hardness values by taking measuremen...

Страница 12: ...6 92 67 57 62 10 6 69 68 93 68 60 65 11 0 71 70 94 69 62 67 11 4 73 72 95 70 64 70 11 8 17 75 74 97 71 67 72 12 2 23 76 75 98 72 69 75 12 6 28 78 77 99 73 72 78 12 9 33 80 79 100 74 75 81 13 3 38 81 8...

Страница 13: ...PCE Instruments 11 93 18 4 103 100 94 18 6 104 101 95 18 7 105 102 96 18 9 106 102 97 19 0 106 103 98 19 2 107 99 19 3 107 100 19 4...

Страница 14: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Страница 15: ...4 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english www pce ins...

Отзывы: