background image

 

 

 

©  PCE Instruments 

 

Troubleshooting 

 
With  the  frequent  use  of  the  tester,  the  pressure  pin  will  be  worn  out  slightly.  When  this 
happens,  the  measurement  value  will  deviate.  So  it  is  necessary  to  check  the  wear  of  the 
pressure  pin  on  a  regular  basis.  If  you  do  not  get  two  readings  which  lie  within  the  indicated 
range  of  the  two  calibration  blocks  when  performing  a  display  calibration,  this  means  that  the 
pressure pin is worn out and needs to be replaced. After that, the device needs to be calibrated 
again. 

7.1 

Pin replacement 

Loosen the screws on both sides of the handle with a small screwdriver (a total of 12). Take off 
the  case  and  take  out  the  spring  tube.  Loosen  the  lower  spindle  connecting  screw  with  a 
spanner and take out the lower spindle. Now you can remove the used pin and replace it with a 
new  one,  then  reinstall  the  lower  spindle  and  lock  it  with  the  spanner.  Put  on  the  spring  tube, 
remount the case and lock it with the screws. See the following picture for reference.  
After replacing the pressure pin, you need to calibrate the device again. 
 

 

 
When the tester is used over a longer period of time, especially when measuring soft aluminium 
alloy  materials,  some  material  might  remain  on  the  pin  which  might  affect  the  accuracy  of  the 
device. To eliminate this error, make sure to check and clean the top of the measuring pin on a 
regular basis. 
 

 

Содержание PCE-1000N

Страница 1: ...N Barcol Hardness Tester User Manual User manuals in various languages Deutsch fran ais italiano espa ol portugu s nederlands t rk polski can be downloaded here www pce instruments com Last change 13...

Страница 2: ...4 3 3 Display 5 4 Getting started 5 4 1 Power supply 5 4 2 Preparation 5 5 Operation 6 5 1 Measurement 6 5 2 Further Functions 6 6 Calibration 7 6 1 Low end calibration 7 6 2 High end calibration 7 6...

Страница 3: ...s The case should only be opened by qualified PCE Instruments personnel Never use the instrument when your hands are wet You must not make any technical changes to the device The appliance should only...

Страница 4: ...ution 0 1 HBa Measurement error 81 88 HBa 1 HBa 42 48 Hba 2 HBa Repeatability error 81 88 HBa 1 5 HBa 42 48 HBa 2 5 Hba Operating conditions temperature humidity 0 50 C 50 RH Power supply 2 x 1 5 V AA...

Страница 5: ...ead The effective measuring range is equivalent to a Brinell hardness of 25 150HB The max hold function shows the maximum value during the measurement The PCE 1000N can be used for measuring not only...

Страница 6: ...0076 HBa Barcol hardness value H indentation depth mm 0 0076 indentation depth at 1 Barcol 2 1 housing screws 2 2 high end calibration button 2 3 low end calibration button 2 4 spring tube 2 5 lower s...

Страница 7: ...lat surface and take a measurement The reading should be in the indicated range of two hardness blocks If it is out of the range a calibration according to chapter 6 is needed The sample surface shoul...

Страница 8: ...asurements taken reaches the set value of the statistical measurement number N Now the device calculates the average value of the measurements taken The result is shown on the right side of the displa...

Страница 9: ...se a small screw driver to loosen the main spindle adjusting screw in counter clockwise direction until the end Put the pin tube into the hole of the pin length calibration block and push the tester s...

Страница 10: ...th a small screwdriver a total of 12 Take off the case and take out the spring tube Loosen the lower spindle connecting screw with a spanner and take out the lower spindle Now you can remove the used...

Страница 11: ...version between different hardness systems is not possible Therefore the following conversion table is for reference only It is recommended to determine the actual hardness values by taking measuremen...

Страница 12: ...6 92 67 57 62 10 6 69 68 93 68 60 65 11 0 71 70 94 69 62 67 11 4 73 72 95 70 64 70 11 8 17 75 74 97 71 67 72 12 2 23 76 75 98 72 69 75 12 6 28 78 77 99 73 72 78 12 9 33 80 79 100 74 75 81 13 3 38 81 8...

Страница 13: ...PCE Instruments 11 93 18 4 103 100 94 18 6 104 101 95 18 7 105 102 96 18 9 106 102 97 19 0 106 103 98 19 2 107 99 19 3 107 100 19 4...

Страница 14: ...d pollutants batteries must not be disposed of as household waste They must be given to collection points designed for that purpose In order to comply with the EU directive 2012 19 EU we take our devi...

Страница 15: ...4 Fax 1 561 320 9176 Fax 44 0 2380 98703 9 Fax 39 0583 974 824 info pce americas com info industrial needs com info pce italia it www pce instruments com us www pce instruments com english www pce ins...

Отзывы: