ERROR DISPLAYS AND TROUBLESHOOTING     

ORP 
LCD 
Display 

ATC 
Display 

DISPLAY
 Mode 

Possible cause(s) 
[Action(s)] 

Any 
Value 

"OVER" 
or 
“UNDER” 

Measure 

a. Temperature outside    –10.0 to +
100.0°C range. 
[Bring solution to a lower/higher
temperature.] 
b. Bad sensor. 
[Replace new electrode module .]

"OVER" 
or 
"UNDER" 

-10.0~
100.0°C

Measure 

ORP value outside    -1100 to +1100 
mV value. 
[Clean or Replace electrode] 

SPECIFICATIONS  

ORP 

Absolute Range 

Resolution 

Accuracy 

-1100 to +1100mV

1 mV 

1 mV ± 1 LSD 

Temperature 

Range

Resolution 

Accuracy

-10.0 to 100.0 °C

0.1 °C 

±0.3 °C 

14.0 to 212.0 

 0.2 

±0.6

 

ORP 
Relative ORP calibration range

–999 to 999 mV

(User programmable)

Input impedance

>10

12

Temperature 
Temperature sensor

 

   Thermistor, 10K@ 25°C 

Temperature unit   

Default in factory (°C or 

)

 

General

 

Power 

Meter :    LR44 x 4 

Calendar :    CR2032 

Battery life 

Meter:   200 Hours 

Calendar: 5 years 

Ambient temperature range

 

  0.0 to 50.0 °C 

Display 

98x64 graphic LCD 

Case 

IP67 water-tight Case 

Weight 

107 g 

Содержание PCE-ORP 3

Страница 1: ...33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Unit 11 Southpoint Business Park Ens...

Страница 2: ...ter Description 3 B PCE ORP 3 Graphical Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING PCE ORP 3 5 A Power ON OFF 5 B Set Date and Time 5 C Relative...

Страница 3: ...in specifications for approximately2 3 hours after low battery indicator appears 1 Shown in the right figure 2 Take off the battery cover 3 Remove all of the old batteries and insert a new set of batt...

Страница 4: ...llar 5 Electrode ATC assembly Electrode cap is not shown B PCE ORP 3 Graphical Display 1 DATE MM DD YY 2 ABSolute or RELative ORP mode indicator 3 HOLD mode indicator 4 TIME HH MM 24H format 5 ORP rea...

Страница 5: ...date and time 7 Erase Mode Can erase 50 set data in non volatile memory B Keypad Operations Key Operation Mode Duration Function All except Hold 2 seconds or more Turns meter on off Measure 0 second...

Страница 6: ...ds or more Enters Erase 50 set non volatile memory mode SCROLL Erase 0 second Exits Erase Mode USING PCE ORP 3 A Power ON OFF Shutdown Press HOLD key for 2 seconds or more The unit will turn on and en...

Страница 7: ...000 mV range D Measure The ORP TEMP can be used to measure absolute or relative ORP values 1 Dip the meter into the test solution in the Measure Mode 2 Press MODE key to select absolute or relative OR...

Страница 8: ...Press SCROLL key over 10 seconds to enter erase mode See the figure below 2 Press STORE key the unit will display DATA CLEAR and erase all data in the non volatile memory 3 Press SCROLL key to leave t...

Страница 9: ...ectrode SPECIFICATIONS ORP Absolute Range Resolution Accuracy 1100 to 1100mV 1 mV 1 mV 1 LSD Temperature Range Resolution Accuracy 10 0 to 100 0 C 0 1 C 0 3 C 14 0 to 212 0 0 2 0 6 ORP Relative ORP ca...

Отзывы: