background image

C.

Relative ORP Calibration

If the user requires a relative ORP calibration, follow these steps: 

1. Clean and dip meter into

standard solution .

2. Wait reading stable,    press

“MODE” key for 2 to 3 seconds
to enter “Calibration” mode and
enter new relative ORP value.

3. Press “HOLD” key to move the

digit position ( polarity & digit).

4. Press “MODE” and “SCROLL”

keys to increase and decrease
values respectively.

5. Press “STORE” key for 2 seconds to exit and return to measure

mode displaying absolute ORP value.

6. Press “MODE” key to display relative ORP value.

Note:

1. The ORP offset calibration is only needed in some applications.
2.

Don’t enter calibration mode when absolute reading is over ±1000

mV range.

D.

Measure

The ORP/TEMP can be used to measure absolute or relative ORP 
values. 

1. Dip the meter into the test solution in the “Measure Mode”.
2. Press “MODE” key to select absolute or relative ORP readings.

E.

Hold Data

1. When the ORP reading is

stable, press “HOLD” key once
to lock the reading.

Note: 

The unit cannot be power 

off in “Hold Mode”. 

2. Press “HOLD” key again to

unlock reading and the unit will
return to “Measure Mode”. The
unit is now ready for another
measurement.

F.

Save Data

1. Either in the “Hold Mode” or the “Measure Mode”, press “STORE”

key for 2-3 seconds. The “ SAVE” and “M-XX” icons will appear
indicating the reading has been saved and stored in memory
location XX. The unit will automatically returns to “Measure Mode”
and ready to perform other task.

Note: 

The 

PCE-

ORP 3

 is equipped with a non-volatile 

memory that can store up to 50 different sets of readings

 

 

6

Содержание PCE-ORP 3

Страница 1: ...33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Unit 11 Southpoint Business Park Ens...

Страница 2: ...ter Description 3 B PCE ORP 3 Graphical Display 3 OPERATION MODES AND KEYPAD OPERATIONS 4 A Operation Modes 4 B Keypad Operations 4 USING PCE ORP 3 5 A Power ON OFF 5 B Set Date and Time 5 C Relative...

Страница 3: ...in specifications for approximately2 3 hours after low battery indicator appears 1 Shown in the right figure 2 Take off the battery cover 3 Remove all of the old batteries and insert a new set of batt...

Страница 4: ...llar 5 Electrode ATC assembly Electrode cap is not shown B PCE ORP 3 Graphical Display 1 DATE MM DD YY 2 ABSolute or RELative ORP mode indicator 3 HOLD mode indicator 4 TIME HH MM 24H format 5 ORP rea...

Страница 5: ...date and time 7 Erase Mode Can erase 50 set data in non volatile memory B Keypad Operations Key Operation Mode Duration Function All except Hold 2 seconds or more Turns meter on off Measure 0 second...

Страница 6: ...ds or more Enters Erase 50 set non volatile memory mode SCROLL Erase 0 second Exits Erase Mode USING PCE ORP 3 A Power ON OFF Shutdown Press HOLD key for 2 seconds or more The unit will turn on and en...

Страница 7: ...000 mV range D Measure The ORP TEMP can be used to measure absolute or relative ORP values 1 Dip the meter into the test solution in the Measure Mode 2 Press MODE key to select absolute or relative OR...

Страница 8: ...Press SCROLL key over 10 seconds to enter erase mode See the figure below 2 Press STORE key the unit will display DATA CLEAR and erase all data in the non volatile memory 3 Press SCROLL key to leave t...

Страница 9: ...ectrode SPECIFICATIONS ORP Absolute Range Resolution Accuracy 1100 to 1100mV 1 mV 1 mV 1 LSD Temperature Range Resolution Accuracy 10 0 to 100 0 C 0 1 C 0 3 C 14 0 to 212 0 0 2 0 6 ORP Relative ORP ca...

Отзывы: