Manual
3
1. Characteristics
-
Measuring range 0 - 20 kg, highest resolution and accuracy
-
Maximum value registration
-
Adjustable memory interval
-
Large LCD display with backlight
-
Measured values can be transmitted to the computer as an Excel file
-
3 display units: kg, lb, N
-
Measuring possibilities for tensile and compressive forces
-
Maximum value can be held on the display
-
Zeroing for all modes possible
-
Display direction: forward / reverse
-
The device can be attached to a test stand (optional)
-
Low battery consumption / mains adaptor optional
-
Battery indicator
-
Microprocessor circuit
-
Overload protection
2. Specifications
Display
5-digit, 16 mm LCD, backlight
Display direction
Selectable via button on the front side
Functions
Tensile and compressive force measurement
Peak hold
Stores maximum value (max. load)
ZERO
For normal and maximum value mode
Measuring range
20 kg / 44.10 lb / 196.10 N
Resolution
0.01 kg / 0.01 lb / 0.02 N
Min. display value
0.02 kg / 0.07 lb / 0.3 N
Accuracy
±(0.
5 % of the display value + 2 digits) at 18 … 28 °C
Units
Kg, N, lb
Sampling rate
Every 0.2 s in quick mode / every 0.6 s in slow mode
Overload protecion
Shows „- - - -“ in case of overload
Max. load
30 kg / 66.10 lb / 294.10 N
Power supply
6 x 1.5 V AA battery or DC 9V adaptor
Operating temperature
0…50 °C
Operating humidity
<80 % RH
Dimensions
215 x 90 x 45 mm
Weight
650 g
Normal measuring
mode
Saving interval
Auto data acquisition
Adjustable, 1 s
… 9 h
(saving interval can be set to 1 second,
but this may lead to data loss)
Manual data registration
Measurement when the recording button
is pressed
(set the saving interval to 0 seconds)
(in manual mode, the memory position
numbers 1-99 can be selected)
Sampling rate in Peak
Hold mode
10 ms …. 500 ms, in steps of 10 mS
Memory capacity
1000 values
Further settings
-
SD card
-
Data / time (year / month / day / hour / minute / second)
-
Saving interval / sampling
-
Auto power off
-
Warning or beeper (on / off)