5.3
Average Mode
In Average (AVG) mode, the average value of a measurement is displayed. There are two
different functions in this measurement mode.
MOD1:
With this function, the average value of the force curve is displayed starting from the set
minimum force and over the set period of time.
MOD2:
This function calculates the average above the set minimum measured value. When the
measured value falls below the set minimum value again, the measurement is finished. This
measuring procedure is possible over a period of 10 minutes. As long as the measuring time of
10 minutes is not exceeded, this measurement can be resumed at any time.