Sensors install and connect according to detail procedures described in specific sensor 
manuals.  Some common installation methods include stud mount, Petro-wax, 
cyanoacrylate adhesive, and magnetic.  The method used depends on test conditions and 
user preference. 
 
In general, sensors connect to the ICP

® 

power unit at the jack labeled "XDCR."  The jack 

labeled "SCOPE" is connected to the readout device (FFT analyzer/computer).  Consult 
ICP

®

 Power Unit Operating Guide for proper installation and operating procedures. 

 

 
4.0 OPERATION 

1. With cables supplied, connect the hammer to a power unit, and then to your analyzer, 
as shown in the system and power unit operation guide. 
 
2. Tighten the cable connectors securely by hand to insure good electrical contact.  
 
3. Switch power on, and wait a minute or two for the sensor amplifier to turn on, and for 
the coupling capacitor to fully charge. 
 
4. Check the power unit meter for normal operation (meter pointer pointing in green 
area).  
 
5. If meter pointer points in the red area, look for shorted cables or connections. If meter 
pointer points in the yellow area, look for open cables or connections. 
 
Connect accelerometer(s) in similar manner referring to the appropriate operating guides 
for the accelerometer(s) and power unit.  When all power unit meters indicate normal 
operation (green), proceed with tests, following all sensor, power unit, and analyzer 
operating instructions. 
 

 
4.1 TESTING 

To test the behavior of your structure, and to tailor the frequency bandwidth of the force, 
follow the following procedures: 
 
1. Strike the test object with the hammer and process the results. Always take several 
averages to reduce the effects of spurious noise.  

CAUTION:

  Never impact without a 

hammer tip properly installed on the sensor element.  In the case of the 086D80, the 
sensor element is pre-installed with a steel tip. 
 
2. Check the measured results for signal quality (adequate signal-to-noise), no overloads 
(overload lights or sharp flattening of time history peaks), and no double impacts. 
 
3. Analyze results for frequency content, and check to insure that the reasonably flat 
portion of the force spectrum is sufficient to cover the structural resonances present in the 
acceleration spectrum.  Often signal energy is sufficient to excite structural resonances at 

Содержание Vibration Division ICP 086C01

Страница 1: ...on and Operating Manual For assistance with the operation of this product contact the Vibration Division of PCB Piezotronics Inc Division toll free 888 684 0015 24 hour SensorLineSM 716 684 0001 Fax 7...

Страница 2: ...id the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment cali...

Страница 3: ...warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and...

Страница 4: ...ion with an accelerometer or the acoustic signature with a microphone Structures generally respond as 1 rigid or elastic bodies 2 finite elements lumped constant models and 3 distributed parameter mod...

Страница 5: ...tive frames tanks and bridges require an instrumented sledgehammer small structures like compressor blades often require mini hammers Some very large structures may require a massive mechanical ram in...

Страница 6: ...3 Models 086C02 086C03 086C04 Model 086D05...

Страница 7: ...4 Model 086D20 Model 086D50...

Страница 8: ...l Many FFT analyzer computer systems have ICP power supply built in The hammer is a single integral unit Laser welded construction of the sensor element insures reliable operation in adverse environme...

Страница 9: ...tions If meter pointer points in the yellow area look for open cables or connections Connect accelerometer s in similar manner referring to the appropriate operating guides for the accelerometer s and...

Страница 10: ...with lower stiffness require the use of the extender and soft impact tip to adequately excite low frequency resonances During testing occasionally check and tighten the electrical and mechanical conne...

Страница 11: ...having as a rigid body is a consistent 1 M ratio the force and the acceleration signals produces a calibration constant ideally 1 M for each discrete frequency The effects of a non modally tuned hamme...

Страница 12: ...y time for any reason you are not completely satisfied with any PCB product PCB will repair replace or exchange it at no charge You may also choose within the warranty period to have your purchase pri...

Страница 13: ......

Страница 14: ......

Отзывы: