 QUARTZ STRAIN SENSOR OPERATION MANUAL 

 

 4 

5.0 POLARITY 

 
Extension  of  the  mounting  area  of  an  ICP  strain  sensor 
produces a positive-going voltage output. The retraction of the 
mounting area produces a negative-going voltage output. 
 

6.0 LOW-FREQUENCY MONITORING 

 
Strain sensors used for applications in short term, steady-state 
monitoring,  such  as  sensor  calibration,  or  short  term,  quasi-
static  testing  should  be  powered  by  signal  conditioners  that 
operate  in  DC-coupled  mode.    PCB  Series  484  Signal 
Conditioner  operates  in  either  AC  or  DC-coupled  mode  and 
may be supplied with gain features or a zero 

clamped

 output 

often  necessary  in  repetitive,  positive  polarity  pulse  train 
applications. 
 
If  you  wish  to  learn  more  about  ICP  sensors,  consult  PCB

General  Signal  Conditioning  Guide,  a  brochure  outlining  the 
technical specifics associated with piezoelectric sensors.  This 
brochure is available from PCB by request, free of charge. 
 

7.0 DISCHARGE TIME CONSTANT 

 
The  discharge  time  constant  (DTC)  of  the  entire  transduction 
system  from  sensor  to  readout  must  be  considered  when 
attempting to calibrate an ICP strain sensor by static methods. 
In order to take full advantage of the long DTC built into the 
strain  sensor,  it  is  best  to  DC  couple  from  the  sensor  to  the 
readout  device.    Several  dual-mode  PCB  signal  conditioners 
(e.g.,  Series  484)  use  direct  coupling  techniques  to  decouple 
the output signal from the sensor bias voltage.  With the output 
of  the  signal  conditioner  coupled  to  a  DC  readout,  such  as  a 
digital  voltmeter  (DVM) or oscilloscope, the time constant of 
the  sensor  is  not  compromised  by  AC  coupling  elsewhere  in 
the system. 
 
When  DC  coupling  to  a  system,  it  is  important  to  DC  couple 
the  entire  system  and  not  just  from  the  sensor  to  the  signal 
conditioner.    The  system  time  constant  is  determined  by  the 
shortest time constant in the system.  For this reason, the signal 
conditioner,  as  well  as  the  readout  device,  must  be  DC 
coupled. 

 

Figure 7.1 Characteristic Discharge Time Constant Curve 

The  discharge  time  constant  represents  the  decay  rate  of  an 
input signal.  One DTC represents the amount of time taken for 
the  signal  to  decay  to  37%  of  the  initial  peak  value.    As 
illustrated  in 

Figure  7.1

,  this  is  an  exponential  decay.  

Approximately five DTC intervals are needed for a peak signal 
to naturally decay back to zero. 
 
The  rule  of  thumb  for  signal  discharge,  as  outlined in 

Figure 

7.2

,  is  this:  for  the  first  10%  of  the  DTC,  the  signal  lost  is 

approximately proportional to the time elapsed. 
 

 

Figure 7.2 Step Function Response 

 
For  example,  a  sensor  with  a  150-second  DTC  loses 
approximately 1% of its output level the first 1.5 seconds (1% 
of  150)  after the application of a steady state strain within the 
measuring range.  In this case, the output reading must be taken 
within 1.5 seconds of the strain application for 1% accuracy. 
 
If  it  is  impossible  to  avoid  AC  coupling  somewhere  in  the 
sensing system, try to keep the coupling DTC at least an order 
of  magnitude  longer  than  the  DTC  of  the  strain  sensor.    This 
avoids compromising the sensor DTC.  
 

8.0  CALIBRATION 
 

Strain sensors are calibrated relative to a strain gage reference 
sensor. A calibration certificate is supplied with each strain 
sensor providing its relative voltage sensitivity (mV/



). A 

calibration must be performed once strain sensors are installed 
in the specific equipment being measured. This is necessary so 
that a direct comparison of relative data can be made thereby 
allowing the user to set control limits and properly monitor a 
specific event as well as the entire process.

 

Содержание RHM240A02

Страница 1: ...ain Sensor Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...model number s purchased The following pages give a brief description of the sensor series recommended mounting procedures operation and recommended calibration In addition to standard products PCB h...

Страница 7: ...ould remain on the connector to prevent contamination or damage For installation in dirty humid or rugged environments it is suggested that the connection be shielded against dust or moisture with shr...

Страница 8: ...ning to all ICP instrumentation Most of the signal conditioners manufactured by PCB have an adjustable current feature allowing a choice of input currents from 2 to 20 mA In general for lowest noise b...

Страница 9: ...ortant to DC couple the entire system and not just from the sensor to the signal conditioner The system time constant is determined by the shortest time constant in the system For this reason the sign...

Страница 10: ...allow this output voltage to decay 5 DTC s prior to making measurements in order to prevent baseline shift where an absolute zero is needed throughout entire monitoring process 10 0 MAINTENANCE The s...

Страница 11: ...ngth x Height 0 67 in x 1 81 in x 0 6 in 17 mm x 46 mm x 15 2 mm Weight 1 6 oz 45 gm 1 All specifications are at room temperature unless otherwise specified In the interest of constant product improve...

Страница 12: ...17 MJK 09 27 17 22493 PCB Piezotronics Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent...

Отзывы: