𝑓

max=

109

2𝜋𝐶𝑉

[𝐿𝑐−1]

    (Equation 1) 

 
Where,  f

max

 = maximum frequency (hertz)     

C = cable capacitance (picofarads)   
V  =  maximum  peak  output  from  sensor 
(volts)     
L

c

  =  constant  current  from  signal 

conditioner (mA)  

 

10

9

 = scaling factor to equate units 

 
Note that in this equation, 1 mA is subtracted from the total 
current supplied to sensor (L

c

). This is done to compensate 

for powering the internal electronics. Also, note that these 
are typical values only.  
 
When  driving  long  cables,  Equation  1  shows  that,  as  the 
length  of  the  cable,  peak  voltage  output  or  maximum 
frequency of interest increases, a greater constant current 
will be required to drive the signal. 
 
The  nomograph  on  page  12  of  this  manual  provides  a 
simple,  graphical  method  for  obtaining  the  expected 
maximum  frequency  capability  of  an  ICP®  measurement 
system. The maximum peak signal voltage amplitude, cable 
capacitance and supplied constant current must be known 
or presumed.  
 
For example, when running at 100 ft (30.5 m) cable with a 
capacitance  of  30  pF/ft,  the  total  capacitance  is  3,000pF. 
This  value  can  be  found  along  the  diagonal  cable 
capacitance  lines.  Assuming  the  sensor  operates  at  a 
maximum output range of 5 volts and the constant current 
available from the power supply is 2 mA, the ratio on the 
vertical axis can be calculated to equal 5. The intersection of 
the  total  cable  capacitance  and  this  ratio  result  in  a 
maximum frequency of approximately 10.2 kHz. 
 
The nomograph does not indicate whether the frequency 
amplitude  response  at  a  point  is  flat,  rising  or  falling.  For 
precautionary  reasons,  it  is  good  general  practice  to 
increase  the  constant  current  (if  possible)  to  the  sensor 
(within  its  maximum  limit)  so  that  the  frequency 
determined from the nomograph is approximately 1.5 to 2 
times greater than the maximum frequency of interest. 

 
Note: 

Higher  current  levels  will  deplete  battery-powered 

signal  conditioners  at  a  faster  rate.  Also,  any  current  not 
used  by  the  cable  goes  directly  to  power  the  internal 
electronics and will create heat. This may cause the sensor to 

exceed  its  maximum  temperature  specification.  For  this 
reason, do not supply excessive current over short cable runs 
or when testing at elevated temperatures.  
 

5.1 CABLE CONNECTOR PROCEDURE 

Care  and  attention  to  installation  is  essential,  as  the 
reliability and accuracy of your system is no better than that 
of the output cable.  

 
STEP 1: 

Ascertain that you have ordered the correct cable 

type.  
 
As with sensors, no cable can satisfy all applications. Special 
low-noise  cabling  should  be  used  with  high  impedance, 
charge-output  devices.  ICP®  sensors  usually  operate  with 
any  ordinary  tow-wire  cable.  Industrial  applications  often 
require twisted/shielded cables to reduce the effects of EMI 
and  RFI.  PTFE-jacketed  cabling  may  be  necessary  to 
withstand corrosive environments.  

 
STEP  2: 

Connect  the  cable  to  the  accelerometer.  A  small 

amount  of  thread-locking  compound  placed  on  the 
connector prior to attachment helps secure the cable during 
testing.  In  harsh  environments,  the  connection  can  be 
sealed with silicone rubber, O-rings, and flexible heat-shrink 
tubing. 

 
STEP  3: 

Plug  the  connector  of  the  cable  into  the  mating 

sensor  connector.  Then,  holding  the  sensor  stationary, 
secure  the  connector  in  place  by  tightening  down  the 
attached threaded cable sleeve.  

 
STEP 4: 

Route the cable to the signal conditioner, making 

certain  to  stress-relieve  the  sensor/cable  connection  and 
minimize motion by clamping the cable at regular intervals.  
 
Common sense should be used to avoid physical damage 
and  minimize  electrical  noise.  For  instance,  avoid  routing 
cables  near high-voltage  wires. Do  not  route  cables  along 
floors or walkways where they be stepped on or become 
contaminated.  Shielded  cables  should  have  the  shield 
grounded at one end only, normally at the instrumentation 
end.  

 
STEP  5: 

Finally,  connect  the  remaining  cable  end  to  the 

signal  conditioner  or  readout  device.  To  dissipate  charge 
that may have accumulated in the cable short the signal to 
ground prior to attachment.  

 
 

Содержание IMI SENSORS EX637 Series

Страница 1: ...el 25mV g 4 to 1 5kHz side exit 2 pin conn Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 800 959 4464 24 hour SensorLin...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...lated 2 20 mA at 18 to 28 VDC source This power is regulated by a current limiting circuit which provides the constant current excitation required for proper operation of ICP sensors In general batter...

Страница 7: ...sential as the reliability and accuracy of any measurement system is no better than that of its weakest link Due to the nature of vibration measurements all sensor cables will ultimately fatigue and f...

Страница 8: ...CB Piezotronics This Division is wholly contained in the PCB Piezotronics manufacturing facility at 3425 Walden Avenue Depew New York Same address same manufacturing facility Some of the documentation...

Страница 9: ...ON A ECN NUMBER 49766 Operating Guide with Enclosed Warranty Information 3424 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 959 4IMI Piezoelectric IC...

Страница 10: ...ply low frequency and high frequency measurements Shear design sensors are preferred because of their inherent insensitivity to adverse environmental influences such as case or base strain and thermal...

Страница 11: ...wer gain which is extremely close to unity and independent of input voltage Also the diode can be changed to supply higher currents for driving long cable lengths Constant current diodes as shown in F...

Страница 12: ...nics However when operating AC coupled signal conditioners with readout devices having an input impedance of less than one megaohm thelow frequencyrangemaybeaffected 4 1 STANDARD STUD MOUNT This mount...

Страница 13: ...derside of the mounting base Firmly press down on the assembly to displace any extra adhesive remaining under the base See Figure7 METHOD 2 Direct Adhesive Mount For restrictions of space or for conve...

Страница 14: ...interference or radio frequency interference EMI or RFI To avoid ground loops thereshouldonlybeonegroundin the system It is recommended for permanent installations that the sensor becase isolatedand...

Страница 15: ...ate Also any current not used by the cable goes directly to power the internal electronicsandwillcreateheat Thismaycausethesensorto exceed its maximum temperature specification For this reason donotsu...

Страница 16: ...er the meter or LED and allow the sensor to power up If a faulty condition is monitored first check all system connections then check the functionality of the cable and signal conditioner If the syste...

Страница 17: ...and former MIL STD 45662A and uses equipment directly traceable to NIST This assures an accurate calibration of relevant specifications The following includes a broad overview of the Back to Back Cal...

Страница 18: ...A of constant current to extend battery life while continuous monitoring systems offer more current in order to drive longer cables Figure 16 shows a typical system for a sensor with integral electron...

Страница 19: ...eive assistance via e mail at imi pcb com or visit our website at www pcb com 11 0 CUSTOMER SERVICE WARRANTY IMI a division of PCB Piezotronics guarantees Total Customer Satisfaction If at any time fo...

Страница 20: ......

Страница 21: ...O DRAFTING 49059 NO CHANGES WITHOUT APPROVAL SCHEDULE DRAWING NO MODIFICATIONS PERMITTED WITHOUT REFERENCE TO THE NOTIFIED BODY SHIELDS TO BE EARTHED AT BARRIER ENDS THE INSTALLER SHALL INSURE THAT TH...

Страница 22: ...n Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Piezotronics Inc REVISIONS REV DESCRIPTION DIN SEE SHEET 1 SCHEDULE DRAWING NO MODIFICATIONS PE...

Страница 23: ...UNF 5 MountingTorque 2 to5 ft lb 2 7 to6 8 Nm SensingElement Quartz Quartz SensingGeometry Shear Shear HousingMaterial Stainless Steel Stainless Steel Sealing Hermetic Hermetic 6 Electrical Connector...

Страница 24: ...CASE ISOLATED ACCELEROMETER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 KRM 12 13 18 KRM 12 13 18 RMB 12 13 18 68326 PCB Piezotronics Inc claims proprietary rights in the information disclosed here...

Страница 25: ...tes that the product is subject to the special conditions of use specified in the Schedule to this certificate 9 This EU Type Examination Certificate relates only to the design and construction of the...

Страница 26: ...The relevant Essential Health and Safety Requirements EHSR5 have been identified and assessed in lntertek Report 103312845CRT 003a Dated 8 August 2019 This Certificate is forth exclusive use of Intert...

Страница 27: ...9 MODEL EX M 637XYYYZ SERIES ETCHING DRAWING Page 1 o 2 This Certificate Is for the exclusive use of Intertek s client and is provided pursuant to the agreement between Intertek and its Client Interte...

Страница 28: ...f Intertek s client and is provided pursuant to the agreement between Intertek and its Client lntertek n responsibility and liability are limited to the terms and conditions of the agreement Intertek...

Страница 29: ...number it indicates that the product is subject to the special conditions of use specified in the Schedule to this certificate 9 This Type Examination Certificate relates only to the design of the spe...

Страница 30: ...C 60079 15 Unused connector must be fitted with an appropriately rated blanking cover b Conditions of Manufacture Routine Tests Dielectric Strength Test Per IEC 60079 15 14 Essential Health and Safety...

Страница 31: ...3 ETCHING DRAWING 70897 NR 8 8 19 MODEL EX M 637XYYYZ SERIES ETCHING DRAWING Page lof 2 This Certificate is for the exclusive use of lutertek s client and is provided pursuant to the agreement between...

Страница 32: ...of lntertek s client and is provided pursuant to the agreement between Intertek and its Client ietertek s responsibility and liability are limited to the terms and conditions of the agreement Intertek...

Страница 33: ...Number This Authorization to Mark is for the exclusive use of Intertek s Client and is provided pursuant to the Certification agreement between Intertek and its Client Intertek s responsibility and li...

Страница 34: ...nd Non Incendive Equipment For Use In Hazardous Locations R2016 CSA C22 2 157 1992 Ed 3 GI U2 Nonincendive Electrical Equipment For Use In Class I And II Division 2 And Class III Divisions I And 2 Haz...

Страница 35: ...YYYZ EX M 638XYYYZ Optional accessory Type of Protection IntrinsicSafety Ia Non Sparking nA Marking Ex ia IIC T4 Ga Ex nA IIC T4 Gc 1960 5 Ta 5 121 C lECEx ETL 19 0034X Approved for issue on behalf of...

Страница 36: ...erational Documents as amended STANDARDS The apparatus and any acceptable variations to it specified in the schedule of this certificate and the identified documents was found to comply with the follo...

Страница 37: ...barrier item When used in Zone 2 The Accelerometers provide a voltage output when subjected to mechanical motion They are intended to be fixed installation in the explosive atmosphere and to be suppl...

Страница 38: ...98 NR 7 24 2019 SECURITE INTRINSEQUE NON SPARK ARC lECEx Ex Ia lECEx Ex nA EX M 637XYYYZ Series Industrial ICP Accelerometer EX M 638XYYYZ Series Industrial ICP Accelerometer Page 2 of 2 DESCRIPTIVE N...

Страница 39: ...ETCHING DRAWING Page 1 of 2 ETCHING DRAWING 70900 NR 8 8 19 MODEL EX M 638XYYYZ SERIES ETCHING DRAWING Page 2 of 2 APPROVAL 70910 NR 7 27 19 INTERCONNECTION DIV 1 ZONE 0 Page 1 of APPROVAL 70910 NR 7...

Страница 40: ...INSTRUCTIONS FOR USE 70911 NR 08 01 19 EX M 637XYYYZ and EX M 638XYYYZ Series Page 3 of 3 APPROVAL XTAL 71165 NR 7 24 19 DRAWING Page 1 of 1 Note An is included before the title of documents that are...

Страница 41: ...data sheets and where those data sheets refer to this declaration of conformity The data sheets for all ICP model numbers referenced above which include the CE mark on such data sheets and refer to th...

Страница 42: ...045c GM29047c GM29029s GM29031s GM29033s GM29044s GM29046s GM29048s I the undersigned hereby declare that the equipment specified above conforms to the above Directive s Standard s Place Depew NY Date...

Страница 43: ...erenced above which include the CE mark on such data sheets and refer to this Declaration of Conformity are hereby incorporated by reference into this Declaration Conform to the following UK Legislati...

Страница 44: ...045c GM29047c GM29029s GM29031s GM29033s GM29044s GM29046s GM29048s I the undersigned hereby declare that the equipment specified above conforms to the above Directive s Standard s Place Depew NY Date...

Отзывы: