background image

 

 

6.0 POWERING

 

All  ICP®  sensors  require  constant  current  excitation 
for  proper  operation.  For  this  reason  only  use 
approved constant-current sources. A typical system 
schematic is shown in Figure 11.  

 

 
The  power  supply  consists  of  a  regulated  18  to  30 
VDC  source.  In  general,  battery-powered  devices 
offer 

versatility 

for 

portable, 

low-noise 

measurements, whereas line-powered units provide 
the capability for continuous monitoring. This power 
is  regulated  by  a  current-limiting  circuit,  which 
provides the constant-current excitation required for 
proper  operation  of  sensors  with  integral 
electronics. 

 
Note: 

Under no circumstances should a voltage be  

supplied to an ICP® accelerometer without a current-
regulating diode or equivalent electrical circuit. 

A  capacitor  at  the  output  stage  of  the  device 
removes  the  sensor  output  bias  voltage  from  the 
measurement signal. This provides a zero-based, AC-
coupled  output  compatible  with  most  standard 
readout devices.  
 
Today,  many  FFT  analyzers,  data  acquisition 
modules,  and  data  collectors  have  constant-current 
excitation for direct use with ICP® sensors. However, 
before  using  this  feature,  check  that  the  supply 
voltage  and  constant  current  are  adequate  for  use 
with  your  sensor.  Please  contact  the  respective 
signal  conditioner  manufacturer  or  check  the 
product manual for more information. 

 
7.0 OPERATING 

After  completing  the  system  setup,  switch  on  the 
conditioner, the meter (or LED) and allow the sensor 
to power up. If a faulty condition is monitored, first 
check  all  system  connections,  then  check  the 
functionality  of  the  cable  and  signal  conditioner.  If 

the  system  still  does  not  operate  properly,  consult 
an IMI Application Engineer. 
 

Note: 

Always  operate  the  accelerometer  within  the 

limitations  listed  on  the  Specification  Sheet  of  your 
sensor.  Operating  the  device  outside  of  its 
parameters  can  cause  temporary  or  permanent 
damage to the sensor. 

 
8.0 ACCELEROMETER CALIBRATION 
Note:

  It  is  good  measurement  practice  to  verify  the 

performance of each accelerometer before and after 
each measurement.  

 
Accelerometers 

are 

precision 

measuring 

instruments.  They  are  highly  engineered  to  provide 
accurate  electrical  signals  representative  of  the 
vibration being monitored. Each sensor is calibrated 
by  comparison  to  a  known  acceleration  level.  Some 
calibrations  include  frequency  response  curves, 
resonant 

frequency 

measurements, 

transvers 

sensitivity,  and  many  other  valuable  tests. 
Applications  requiring  a  high  degree  of  accuracy  or 
those in plants requiring certification and traceability 
require full calibration test results.  

 

 
Some  applications  have  much  less  stringent 
requirements  for  calibration  certification.  Simple 
sensitivity  measurements  at a  single  frequency  may 
be sufficient; operational verification and certificates 
of  conformance  to  published  specifications  may 

satisfy many plants’ calibration needs. Reducing the 

final  calibration  requirements  reduces  the  cost  of 
manufacturing the sensor and should lower the price 
for predictive maintenance users. 

Содержание IMI SENSORS 602D11/015AC

Страница 1: ...Accelerometer Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 959 4464 24 hour SensorLine 716 684 0001 Fax 716 684 3823...

Страница 2: ...ty regimen When in doubt about a calibration cycle a good rule of thumb is to recalibrate on an annual basis It is also good practice to recalibrate after exposure to any severe temperature extreme sh...

Страница 3: ...rchase Order should include authorization to proceed and return at current pricing which can be obtained from a factory customer service representative Contact Information International customers shou...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...ON A ECN NUMBER 49766 Operating Guide with Enclosed Warranty Information 3424 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 959 4IMI Piezoelectric IC...

Страница 7: ...ply low frequency and high frequency measurements Shear design sensors are preferred because of their inherent insensitivity to adverse environmental influences such as case or base strain and thermal...

Страница 8: ...wer gain which is extremely close to unity and independent of input voltage Also the diode can be changed to supply higher currents for driving long cable lengths Constant current diodes as shown in F...

Страница 9: ...nics However when operating AC coupled signal conditioners with readout devices having an input impedance of less than one megaohm thelow frequencyrangemaybeaffected 4 1 STANDARD STUD MOUNT This mount...

Страница 10: ...derside of the mounting base Firmly press down on the assembly to displace any extra adhesive remaining under the base See Figure7 METHOD 2 Direct Adhesive Mount For restrictions of space or for conve...

Страница 11: ...interference or radio frequency interference EMI or RFI To avoid ground loops thereshouldonlybeonegroundin the system It is recommended for permanent installations that the sensor becase isolatedand...

Страница 12: ...ate Also any current not used by the cable goes directly to power the internal electronicsandwillcreateheat Thismaycausethesensorto exceed its maximum temperature specification For this reason donotsu...

Страница 13: ...er the meter or LED and allow the sensor to power up If a faulty condition is monitored first check all system connections then check the functionality of the cable and signal conditioner If the syste...

Страница 14: ...and former MIL STD 45662A and uses equipment directly traceable to NIST This assures an accurate calibration of relevant specifications The following includes a broad overview of the Back to Back Cal...

Страница 15: ...A of constant current to extend battery life while continuous monitoring systems offer more current in order to drive longer cables Figure 16 shows a typical system for a sensor with integral electron...

Страница 16: ...eive assistance via e mail at imi pcb com or visit our website at www pcb com 11 0 CUSTOMER SERVICE WARRANTY IMI a division of PCB Piezotronics guarantees Total Customer Satisfaction If at any time fo...

Страница 17: ......

Страница 18: ......

Страница 19: ......

Отзывы: