MODELS 483C30, 483C50, 482M179 AND 483M217 GENERAL OPERATION MANUAL
24
LEDS
The LEDS command is used to test the LED functionality of the front panel. When sent as a command the
LED’s on the front panel will flash 3 times. This command invokes a function and therefore has no query
capability.
Setting:
2:1:LEDS= 0\r\n ( unit 2, channel 1, Any value )
Setting Response: 2:LEDS:ok
Query:
N/A
RSET
The RSET command is used to restore the factory default channel settings. This command invokes a function
and therefore has no query capability.
Setting:
2:1:RSET = 0\r\n
( unit 2, channel 1, Any value )
Setting Response: 2:RSET:ok
Query:
N/A
The factory Defaults are:
•
Gain= 1.0
•
Sensitivity=10.0
•
Full Scale input= 1000.0
•
Full Scale output= 10.0
•
Input Mode = ICP
•
Current excitation=4.0mA
•
Output Filter=Disabled
•
Internal Reference Oscillator=OFF
AUTR
The AUTR command invokes the Auto-Scaling function. With this command, the programmable gain is adjusted to give
the largest signal amplitude output without causing an overload condition. The signals are checked internally; if there is
an overload, the signal is stepped down and checked again.
Possible settings are: 0=off,1-on,2=immediate. The Immediate option causes the auto scale function to execute one time
and then turns off the function automatically. When Auto scale is ‘ON’ the unit will continue to invoke the function until
the function is turned OFF with a subsequent command.
Setting:
2:1:AUTR =1\r\n
( unit 2, channel 1, Auto Scaling ON )
Setting Response: 2:AUTR:ok
Query: