MODELS 483C30, 483C50, 482M179 AND 483M217 GENERAL OPERATION MANUAL

 

16 

 
Command type definitions;  

 

R/W – the setting can be read from or written to the unit or channel. 

 

R – The information can only be read from the unit or channel. 

 

FCN – The command invokes a function in the unit. 

 

3-6. 

Command Format 

 
The communication protocol incorporates the concept of 'Directed' and 'Global' commands at both the Unit and Channel 
level with the following characteristics; 

 

Unit or Channel numbers =0 are global commands that affect either all units or all channels of a particular unit 
or both. 

 

Directed commands that set a unit parameter are always acknowledged (ACK) with an ASCII message that 
indicates '<Unit#>:<CMD>:ok’ if implemented with no errors or  NAK with; <Unit#>:<CMD>:=<error#>’ if 
an error was encountered.   

 

Directed commands that request a particular parameters setting (query) result in a query response being 
returned 

 

No response is ever given to a Global Unit command. 

 

Command Format
 

'Unit#:Ch#:Cmd[=|?]{<value1 >{,< value2 >}}{;Ch#:Cmd[=|?]{<value1>{,< value2>}}}\n\r 

 

 

Each message must be preceded by a Unit# & Channel# (both of which could be 0) 

 

Messages may contain multiple commands separated by a semicolon ‘;’. 

 

The second and subsequent commands in a message shall not contain a unit number but shall contain a channel 
number. 

 

Each command in a message will evoke a response message if one is warranted (not global) 

 

Query’s ('?') can only be directed to one unit but if the channel=0 then each channel's setting will be returned in 
the order 0-MAXCHANNELS separated by a ':' 

 
Command examples

1:0:GAIN=100.2\r\n 

 

 

unit 1,all channels gain set to 100.2 

1:1:GAIN=100.2;2:GAIN=120.3\r\n  

unit 1, channel 1 gain = 100.2; channel 2 gain = 120.3 

1:3:GAIN=100.2;0:FLTR=1\r\n  

 

unit 1, channel 3 gain = 100.2; all channel’s filter = ON 

 
General Query Response Format 
Unit#:Cmd:Ch#=<value>{; Ch#=<value>}…>{; Ch#=<value>}…\r\n 
 
Responses to a query with a channel number=0 will return the setting of each channel in a list separated by 
semicolons ‘;’. 
If the target of the query is a unit setting (ex. current excitation) then the channel number returned is the 1

st

 

channel of the board that processed the command. 

 

3-7. 

Multiple Board Models 

All of these models have multiple main boards installed which will appear as a contiguous set of channels, 
when in fact, physically, they are completely separate and operate somewhat independently. This presents some 
interesting communication challenges. For instance, a global channel command to set 8 channels on a unit will 
be handled by each internal board simultaneously. However only the board with the channels designated as 1-4 
will ACK the command, the other board will remain silent. 
 

Содержание ICP 482M179

Страница 1: ...onditioner Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Страница 2: ...repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps...

Страница 3: ...ry customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...NDITIONER Models 483C30 483C50 482M179 and 483M217 GENERAL OPERATION MANUAL For powering ICP and charge sensors these signal conditioners provide an effective method for managing large numbers of sens...

Страница 7: ...2 1 Sensor Excitation 4 2 2 Input Protection 4 2 3 Input Fault Detection 5 2 4 Input Interface 5 2 5 Gain 5 2 5 1 Programmable Gain 6 2 6 Normalized Output Sensitivity 6 2 7 Overload Detection and Au...

Страница 8: ...Commands 17 GAIN 17 SENS 17 FSCI 18 FSCO 18 INPT 18 IEXC 19 OFLT 20 OSCL 20 RTED 21 ALLC 21 RBIA 22 STUS 22 UNIT 22 UNID 23 LEDS 24 RSET 24 AUTR 24 SAVS 25 WTED 25 Communication Responses 26 Table of...

Страница 9: ...MODELS 483C30 483C50 AND 482M179 GENERAL OPERATION MANUAL iii Figure 5 Selectable Gain Amplifier Configuration 5 Figure 6 Autoscale Overload Window Comparator 7...

Страница 10: ...service by qualified personnel who recognize shock hazards and are familiar with the safety precautions required to avoid injury WARNING 2 This equipment is designed with user safety in mind however...

Страница 11: ...volatile memory Ethernet interface Normalized Output Voltage sensor input signal mode Capable of reading TEDS Model 483C50 includes the following features ICP with variable 2 to 20 mA source current...

Страница 12: ...ignal ground lines of all equipment are tied together The signal grounds of the channels are typically tied together via the case of the input and output BNC connectors The individual channels of thes...

Страница 13: ...n occurs as the constant current of all channels are set These models allow the constant current to be adjusted between 2 and 20 mA when not in isolation mode to provide the required excitation for mo...

Страница 14: ...o reference voltages Vref representing thresholds for short and open When the sensor s bias voltage Vbias exceeds the comparator range the front panel input fault LED lights NOTE Red LED implies input...

Страница 15: ...Some users request the ability calculate the gain using a known input signal along with the sensor sensitivity and desired output level The result of the request impacts the formula Gain 1 Volt Sensi...

Страница 16: ...ed The Vref is equal to the default overload value of 10 volts When the input voltage Vsignal of Figure 6 to the window comparator exceeds the reference voltage limits overload has occurred and the co...

Страница 17: ...d and returns the data in an ASCII Hex representation 2 11 Filtering Most Model 482C s can be ordered with optional Input and or Output Filters which may be enabled and disabled for each individual ch...

Страница 18: ...ass 100 kHz 477M03 Low Pass 50 Hz The user may also define a specific cutoff frequency for the Input Filters that satisfies a specific need Contact the factory regarding individual requirements and mo...

Страница 19: ...provided on the PCB MCSC application software installer CD and available for download at the Lantronix website http www lantronix com device networking utilities tools device installer html Once insta...

Страница 20: ...ets with the proper IP address and ensure the packet payload contains the correct Unit Id in the command header The XPort device should be configured properly for communication with the signal conditi...

Страница 21: ...t the parameters are shown as below and pay particular attention to the Active Connection and Endpoint Port items They should be set as shown below Next check the Serial Settings They should be 19200...

Страница 22: ...MODELS 483C30 483C50 482M179 AND 483M217 GENERAL OPERATION MANUAL 13...

Страница 23: ...MODELS 483C30 483C50 482M179 AND 483M217 GENERAL OPERATION MANUAL 14...

Страница 24: ...set Some commands may not be valid in all units The commands are sent and received from to the host computer in ASCII text format Command Type Scope Meaning GAIN R W Channel Set or read gain SENS R W...

Страница 25: ...t contain a unit number but shall contain a channel number Each command in a message will evoke a response message if one is warranted not global Query s can only be directed to one unit but if the ch...

Страница 26: ...directly the unit will adjust the FSI parameter of the gain equation using the following equation FSI FSO 1000 Gain Sens to ensure it remains valid Setting Response 1 GAIN ok Query The Gain query retu...

Страница 27: ...00 0 Global FSCI Query 1 0 FSCI Global FSCI Response 1 FSCI 1 1000 0 2 1000 0 3 1000 0 4 1000 0 FSCO The FSCO command provides a scaling mechanism to automatically set the gain based on a known input...

Страница 28: ...t set to this value on 482C50 and 482M179 models As a result if the channel was in Voltage mode and this value gets set 0 then the channel s input mode will revert to ICP mode Likewise if the channel...

Страница 29: ...When either the 100 or 1k Hz setting is selected the channel is forced into a charge mode if it isn t set that way already and an internal 100pC signal is generated This signal is then converted to a...

Страница 30: ...fies that the status of the Application register 1 if it contains the basic TEDS data or 0 if not This byte is followed by up to 40 bytes in ASCII Hex format 8 bytes of Application register content if...

Страница 31: ...t Cmd Ch unit status bit map CH1 status bit map CHn status bit map Query 1 1 STUS Query Response 1 STUS 1 0 1 5 5 5 Where Unit Bit 0 1 BAD EEPROM read for channel settings on power up Unit Bit 1 1 BAD...

Страница 32: ...Future Use Currently undefined and available for expansion UNID The UNID command is used to set the units ID number The Unit Id number is critical to remote communications since it indicates to the u...

Страница 33: ...n 1 0 Sensitivity 10 0 Full Scale input 1000 0 Full Scale output 10 0 Input Mode ICP Current excitation 4 0mA Output Filter Disabled Internal Reference Oscillator OFF AUTR The AUTR command invokes the...

Страница 34: ...o the signal conditioner determines what chip it is communicating with and whether application register operations are possible The user does however need to construct the command appropriately to ind...

Страница 35: ...rors are indicated with negative numbers The unit may return one of the following Unit Cmd String OK CR LF Signifies that the last command was entered in the correct format and was performed properly...

Страница 36: ...5 0 V rms 75 0 V rms 2 Spectral Noise 1 Hz 20 0 V Hz 20 0 V Hz 2 Spectral Noise 10 Hz 1 5 V Hz 1 5 V Hz 2 Spectral Noise 100 Hz 1 0 V Hz 1 0 V Hz 2 Spectral Noise 1 kHz 1 0 V Hz 1 0 V Hz 2 Spectral No...

Страница 37: ...NG 60878 1 OF 1 75X MODEL 483C50 SENSOR SIGNAL CONDITIONER DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 BB 3 27 15 ECB 3 27 15 CPH 3 27 15 60878 PCB Piezotronics Inc claims proprietary rights in the...

Отзывы: