SERIES 8162 STRAIN GAGE SIGNAL CONDITIONER MANUAL 

 

 
Either  5-volt  or  10-volt  bridge  excitation  can  be 
selected.    In  general,  5-volt  excitation  is  used  with 
120

  transducers,  and  the  10-volt  excitation  is  used 

with 350

 and 700

 transducers.   

 
Excitation  voltage  is  set  by  the  position  of  the  P1 
Jumper.  See Figure 4.  Jumping pins 1 and 2 will set 
the  conditioner  to  10  V

EXC

.    Jumping  pins  2  and  3 

will set the conditioner to 5 V

EXC

.   

 
6.0

 

Calibration  

 
The Series 8162 can be calibrated using (2) different 
methods.   The first method uses actual or simulated 
Zero  and  Full  Scale  (FS)  inputs.    Actual  inputs  may 
come  from  a  strain  gage  transducer  loaded  with 
calibrated weights.  Simulated inputs may come from 
a strain gage bridge simulator.   
 
The  second  method  utilizes  an  R

CAL

  resistor  to 

simulate  a  load  on  a  strain  gage  bridge.    The  R

CAL

 

resistor,  located  in  the  R12  position  in  Figure  4,  is 
shunted  across  one  arm  of  the  strain  gage  bridge 
when  the  R

CAL

  Switch  (SW1)  is  depressed.    The 

shunt  produces  an  imbalance  equivalent  to  that  of  a 
particular  value  of  mechanical  input.    If  this 
Equivalent Input value is accurately known, it can be 
used as a reference point  for shunt calibration of the 
system.    Upon  completion  of  installation  of  the 
transducer and its associated cabling, the user can: 
 
a.

 

Perform an overall deadweight calibration using 
a  precisely  known  value  of  mechanical  input. 
The  calibration  can  then  be  transferred  to  the 
installed  calibration  resistor  for  convenience  in 
checking later. 

b.

 

Replace  the  installed  calibration  resistor  with 
one  supplied  by  the  transducer  manufacturer,  or 
an  equivalent  resistance  value,  to  achieve  a 
precisely  known  equivalent  input,  allowing  the 
instrument sensitivity to be adjusted correctly. 

c.

 

A 30K ohm resistor is installed at the factory and 
is  used  during  final  checkout  of  the  instrument.  
It  is  replaced  with  a  more  proper  value  resistor 
when  calibrated  with  a  specific  transducer  to 
produce  a  more  accurate  shunt  value.    The 
resistor  is  mounted  internally  on  terminal  R12 
located  on  the  printed  circuit  board  in  the 
instrument.   See Figure 4.  It can be reached by 
removing the cover.  

 

Note: Soldering is not required to remove or install 
the shunt calibration resistor.  When an amplifier is 
purchased  with  a  specific  sensor  as  a  system,  PCB 
will  install  the  proper  value  shunt  resistor  in  the 
amplifier.  
 
R

CAL

 Polarity (+R

CAL 

/ -R

CAL

The  default  R

CAL

  polarity  setting  of  the  Series  8162 

Signal Conditioner is positive (+).  Depending on the 
particular 

sensor 

being 

calibrated 

with 

the 

conditioner,  +R

CAL 

and/or  -R

CAL

  data  is  provided  on 

the calibration certificate. 
 
The  polarity  of  the  R

CAL

  may  be  switched  by 

changing  the  solder  jumper  position  on  the  R

CAL

 

solder tabs located on the internal circuit board of the 
conditioner.    See  Figure  4  for  the  R

CAL

  solder  tab 

locations.    Solder  applied  to  jumper  1  will  yield  a 
+R

CAL

, solder applied to jumper 2 will yield a -R

CAL

.  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Содержание 8162-011A

Страница 1: ...ignal Condition Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987...

Страница 2: ...repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps...

Страница 3: ...ry customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period...

Страница 4: ...isclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O O X O O O O O...

Страница 5: ...e O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indica...

Страница 6: ...er Cabling 2 4 0 Analog Outputs 2 5 0 Transducer Excitation 2 6 0 Calibration 3 7 0 Gain Switches 4 8 0 Mounting 4 Figures 1 Series 8162 Signal Conditioner 2 2 Mounting Hole Dimensions 4 3 Exterior Di...

Страница 7: ...terminal J1 Reference Figure 4 for the J1 terminal location Tighten the screw to fasten the wire in place d Next insert the Ground wire into the screw 2 position on terminal J1 Tighten the screw to f...

Страница 8: ...alue of mechanical input The calibration can then be transferred to the installed calibration resistor for convenience in checking later b Replace the installed calibration resistor with one supplied...

Страница 9: ...t Sensitivity mV V Vexc 5 VDC Sensitivity mV V Vexc 10 VDC SW2 Settings 1 2 3 4 SW2 1 SW ON 7 0 to 11 0 3 5 to 5 5 0 0 0 1 4 6 to 7 0 2 3 to 3 5 0 0 1 0 3 0 to 4 6 1 5 to 2 3 0 1 0 0 2 0 to 3 0 1 0 to...

Страница 10: ...nside View J1 Input Power J3 Output Connections P2 Voltage Output P1 Bridge Excitation SW1 RCAL Switch SW2 Gain Switch 1 RCAL R5 4mA adjust 20mA adjust R22 Fine Span R23 Coarse Zero J2 Transducer Conn...

Страница 11: ...56 6 mm x 94 0 mm x 64 8 mm Electrical Connector input output screw terminal screw terminal All specifications are at room temperature unless otherwise specified In the interest of constant product i...

Страница 12: ...NE DRAWING 27193 1 OF 1 FULL MODEL 8162 SIGNAL CONDITIONER SAC 05 24 13 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 27193 PCB Load Torque Inc claims proprietary rights in the information disclosed h...

Отзывы: