MULTICHANNEL SIGNAL CONDITIONER MODEL 482C27 GENERAL OPERATION MANUAL
46
ground
out
sensor
SENSE +
SENSE -
V EXC -
V EXC +
SIG
GND
SIG -
(internally grounded in RSE mode)
+Vpower
R SHUNT
SIG +
5-3.
Quarter Bridge Mode
Figure 9 Recommended Connections for Quarter Bridge Type Sensors
(sense lead wired for optimum performance)
5-4.
RSE / NRSE Mode for 3 Wire Sensors
Three wire sensors having a power connection, a ground connection, and a voltage output may be connected as shown in
Figure 11 below. The channel should be set to RSE mode which internally grounds the SIG- input. The voltage output
from the sensor (or other source) may then be applied to the SIG+ (non-inverting) input. Set V EXC to unipolar at a
voltage appropriate to power the sensor, and ensure that the V EXC lines are tied to the respective SENSE lines.
If an inversion is desired, use NRSE mode which internally connects SIG+ to ground. The voltage output from the
sensor may then be applied to the SIG- (inverting) input.
Figure 10 Recommended Connections for 3 Wire Type Sensor
(channel in RSE mode)
SIG
GND
V EXC +
internal three
quarter bridge
completion
quarter
bridge
sensor
SIG -
SENSE-
V EXC -
V EXC +
SIG -
SIG +
optional
shunt resistor
(external)
R SHUNT
SENSE +