![PCB Piezotronics 482C27 Скачать руководство пользователя страница 43](http://html1.mh-extra.com/html/pcb-piezotronics/482c27/482c27_general-operations-manual_1514989043.webp)
MULTICHANNEL SIGNAL CONDITIONER MODEL 482C27 GENERAL OPERATION MANUAL
35
1:1:VEXC= -10.00 \r\n
(unit 1, channel 1, sets
minus (-) Bridge Excitation and plus (+)Bridge
Excitation to 10.0 volts
)
1:1:VEXC= 10.0 0\r\n
(unit 1, channel 1, sets
minus (-) Bridge Excitation to 0 and plus (+)Bridge
Excitation to 10.0 volts
)
Setting Response: 1:VEXC:ok
Query
:
The VEXC query returns the voltage excitation value.
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <VEXC>;
Query: 1:1:VEXC?
Response: 1:VEXC:1=-10.00;
Global Query: 1:0: VEXC?
Global Response: 1:VEXC:1=-10.00;2=10.00;3=0.00;4=0.00;
FLTR
The FLTR command enables or disables the Input Filter.
The Input Filter value is sent as an integer value of either 0 -Disable or 1-Enable.
Setting
:
2:1:FLTR= 1\r\n ( unit 2, channel 1, Input Filter Enabled )
Setting Response: 2: FLTR:ok
Query
:
The FLTR query returns the channels Input Filter selection
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <0|1>;
Query: 2:1: FLTR?
Response: 2:FLTR:1=1;
Global Query: 1:0: FLTR?
Global Response: 1:FLTR:1=1;2=0;3=0;4=0;
OFLT
The OFLT command enables or disables the Output Filter.
The Output Filter value is sent as an integer value of either 0 -Disable or 1-Enable.
Setting
:
2:1:OFLT= 1\r\n ( unit 2, channel 1, Output Filter Enabled )
Setting Response: 2: OFLT:ok
Query
:
The FLTR query returns the channels Output Filter.
Query Format: Unit#:Ch#:CMD?
Response format: Unit#:Cmd:Ch#:= <0|1>;
Query: 2:1: OFLT?
Response: 2: OFLT:1=1;
Global Query: 1:0: OFLT?