MULTICHANNEL SIGNAL CONDITIONER MODEL 482C27 GENERAL OPERATION MANUAL
42
2:0:RSET = 1\r\n
( unit 2, channel 0, RSET cmd, 1(TRUE) )
Setting Response: 2:RSET:ok
Query
:
N/A
The factory Defaults are:
Gain = 1.0
Sensitivity = 10.0
Full Scale Input = 1000.0
Full Scale Output = 10.0
Input Mode = ICP
Input Filter = Disabled
Current Excitation = 4.0mA
Voltage Excitation = 0 volts
Output Filter = Disabled
Coupling = AC
Clam p= OFF
Calibration = OFF
Switched Output = OFF
Gain Type = Gain
AUTR
The AUTR command invokes the Auto-Range function. This function (also referred to as auto-scale) provides an
automatic scaling of the output signal by adjusting the programmable gain to give .8 of the Full Scale Output setting.
The signals are checked internally using the onboard channel output A/D. Possible settings are:
0=off, 1-on,
2=immediate
. The Immediate option causes the auto scale function to execute one time and then turns off the function
automatically. When Auto scale is ‘ON’ the unit will continue to invoke the function, adjusting the gain for the current
input level, until the function is turned OFF with a subsequent command.
Setting
:
2:1:AUTR =1\r\n
( unit 2, channel 1, Auto Scaling ON )
Setting Response: 2:AUTR:ok
Query:
Query Format: Unit#:Ch#: AUTR?
Response format: Unit#:Cmd:Ch#:=current state (0=off,1-on,2=immediate)
Query: 2:1:AUTR?
Query Response: 2:AUTR:1=0;
SAVS
The SAVS command is used to store the current channel setting as the default settings that will be restored on power up.
This command is available primarily for units without a front panel display and keypad. Units with a soft key power
button will save the channel settings automatically at power down because the units firmware handles the power button
processing. Units without the soft key have power removed abruptly and don’t not have the opportunity to save the
settings automatically. This command invokes a function and therefore has no query capability.
Setting
:
2:1:SAVS = 1\r\n ( unit 2, channel 1, Any value )