background image

 

 

If the microphone is ever dropped (the effects of this can be minimized by PCB’s rugged grid cap or utilizing 
rugged microphone kits), or comes into contact with any contamination, or is exposed to extreme tempera-
tures, immediate recalibration of the microphone should be performed. When using a rugged grid cap, a 
handheld calibrator with a 1” diameter (example CAL250,  or pistonphone) is required for recalibration, 
along with the larger (079A59) calibration adapter.  Please note that heat from your hands can affect the 
calibration results. It is always good practice to wait a minimum of 30 seconds after placing the microphone 
on a test system before taking a calibration reading in order to minimize the effect that heat from your hands 
has  on  the  calibration.  Changes  in  the  environmental  conditions:  temperature,  humidity  or  atmospheric 
pressure, may cause a change in your microphones sensitivity and thus, changes in your output accuracy. It 
is always recommended to calibrate both before and after your test, within the test environment. Performing 
a single point handheld calibrator check, both before and after your test, ensures reliability of the test data. 
With proper maintenance, the microphone and preamplifier should provide stable and accurate results for 
years to come.  

7.0

 

Calibration 

PCB offers recalibration services for our precision microphones, as well as units produced by other A2LA 
manufacturers.  Our  internal  metrology  laboratory  is  certified  to  ISO  9001,  accredited  by  A2LA  to 
ANSI/IEC 17025 and ANSI/NCSL Z540-1, complies with ISO 10012-1 (and former MIL-STD-45662A), 
and uses equipment directly traceable to one or more of the following National Labs (NIST, PTB or DFM). 
Our investment in equipment, traceability and conformance to industry standards ensures accurate calibra-
tion against relevant specifications. We also carry a line of acoustic calibrators that can be used to calibrate 
microphone sensitivity on site as needed. 

Hand calibration of sensitivity for most prepolarized microphones may be performed with a Larson Davis 
CAL250  or  CAL200  (or  equivalent).    The  CAL250  produces  114  dB  SPL  at  251.2  Hz.    The  CAL  200 
produces either 94 dB or 114 dB (user selectable) at 1000 Hz.  Note that both the CAL250 and CAL 200 
are compliant with ANSI S1.40-2006, Specifications and Verification Procedures for Sound Calibrators, 
Class1 and IEC 60942-2003, Class 1, Sound Calibrators. 

8.0

 

Warranty and Service 

All equipment and repair services provided by PCB Piezotronics, Inc. are covered by a warranty against 
defective material and workmanship under a 

Total Customer Satisfaction

 policy. If, at any time, for any 

reason, you are not completely satisfied with any PCB product, PCB will repair, replace, or exchange it at 
no charge. You may also choose to have your purchase price refunded. See the supplemental sheet, con-
tained with this manual, for information on our service, repair and return policies, procedures and instruc-
tions. When unexpected problems arise, call our 24-Hour SensorLine

SM

 (716-684-0001) to speak with an 

Application Engineer. Visit 

www.pcb.com

 for a complete statement of our warranty. 

 
 
 
 
3425 Walden Avenue, Depew, NY 14043-2495 

24-hour SensorLine

SM

: 716-684-0001 

E-Mail: 

[email protected]

 

U.S.A. Fax: 716-684-0987 

Website: 

www.pcb.com

 

Toll-free (in the U.S.A.): 800-828-8840

 

 
A PCB GROUP COMPANY  

ISO 9001 CERTIFIED 

A2LA ACCREDITED to ISO 17025 

 2014 PCB Group, Inc. In the interest of constant product improvement, specifications are subject to change without notice. 

PCB and ICP are registered trademarks of PCB Group, Inc. Sensor Line is a service mark of PCB Group, Inc. All other trademarks 
are properties of the respective owners. 
 Manual Number: 27042 

Manual Revision: F 
ECO 51020 

 
 

Printed in the U.S.A. 

Содержание 377C41

Страница 1: ...Microphone Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E m...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...in which the instrument will be used Whether for legal purposes or for quality assurance programs these standards will help determine the required quality accuracy and consistency of the instrument St...

Страница 7: ...owered with a 24V or a 12V phantom power supply but this will limit the maximum output voltage Externally Polarized Microphone System Prepolarized Microphone System Phantom Powered Microphone System T...

Страница 8: ...em such as the preamplifier signal conditioner or instrumentation Conversely the Cartridge Thermal Noise CTN specification provides the lowest measurable sound pressure level that can be detected abov...

Страница 9: ...rugged grid cap shown below and as an independent part of a rugged kit to extend the lifespan of the precision measuring device caution should be taken to ensure the delicate diaphragm sensing element...

Страница 10: ...ustry standards ensures accurate calibra tion against relevant specifications We also carry a line of acoustic calibrators that can be used to calibrate microphone sensitivity on site as needed Hand c...

Страница 11: ...Pa 62 dB re 20 Pa 1 Influence of Magnetic Field 60 Hz 80 A m 0 dB re 20 Pa 0 dB re 20 Pa 1 Long Term Stability 68 F 20 C 1000 years dB 1000 years dB Electrical Capacitance Polarized 15 pF 15 pF 1 Pola...

Страница 12: ......

Отзывы: