background image

Model Number

208A11

ICP® FORCE SENSOR

Revision: K
ECN #: 45224

Performance

ENGLISH

SI 

Sensitivity(± 15 %)

500 mV/lb

112,404 mV/kN

Measurement Range(Compression)

10 lb

0.04448 kN

Measurement Range(Tension)

10 lb

0.04448 kN

Maximum Static Force(Compression)

60 lb

0.270 kN

Maximum Static Force(Tension)

60 lb

0.270 kN

Broadband Resolution(1 to 10,000 Hz)

0.0001 lb-rms

0.0004448 N-rms

[2]

Low Frequency Response(-5 %)

0.01 Hz

0.01 Hz

[3]

Upper Frequency Limit

36,000 Hz

36,000 Hz

[4]

Non-Linearity

 1 % FS

 1 % FS

[5]

Environmental

Temperature Range

-65 to +250 °F

-54 to +121 °C

Temperature Coefficient of Sensitivity

 0.05 %/°F

 0.09 %/°C

Electrical

Discharge Time Constant(at room temp)

 50 sec

 50 sec

Excitation Voltage

18 to 30 VDC

18 to 30 VDC

Constant Current Excitation

2 to 20 mA

2 to 20 mA

Output Impedance

 100 Ohm

 100 Ohm

Output Bias Voltage

8 to 12 VDC

8 to 12 VDC

[1]

Output Polarity(Compression)

Positive

Positive

Physical

Stiffness

6 lb/µin

1.05 kN/µm

[2]

Size (Hex x Height)

0.625 in x 1.17 in

15.88 mm x 29.7 mm

Weight

0.80 oz

22.7 gm

Housing Material

Stainless Steel

Stainless Steel

Sealing

Epoxy

Epoxy

Electrical Connector

10-32 Coaxial Jack

10-32 Coaxial Jack

Electrical Connection Position

Axial

Axial

All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.

ICP

®

 is a registered trademark of PCB Group, Inc. 

N

- Negative Output Polarity

Output Polarity(Compression)

Negative

Negative

W

- Water Resistant Cable

Electrical Connector

Sealed Cable

Sealed Cable

NOTES:

[1]Sensor contains protected electronics which may cause long 'turn-on" times.
[2]Typical.

[3]Calculated from discharge time constant.
[4]Estimated using rigid body dynamics calculations.
[5]Zero-based, least-squares, straight line method.
[6]See PCB Declaration of Conformance PS023 for details.

SUPPLIED ACCESSORIES: 

Model 080A81 Thread Locker (1)
Model 081B05 Mounting Stud (10-32 to 10-32) (1)

Model 084A03 Impact Cap (1)
Model M081A62 Mounting stud, 10-32 to M6 x 1, BeCu with shoulder (1)

3425 Walden Avenue, Depew, NY 14043

Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]

[6]

OPTIONAL VERSIONS

Optional versions have identical specifications and accessories as listed for the standard model 

except where noted below. More than one option may be used.

Entered: LK

Engineer: MJK

Sales: KWW

Approved: APB

Spec Number:

Date: 3/30/2016

Date: 3/30/2016

Date: 3/30/2016

Date: 3/30/2016

208-1110-80 

Содержание 208A11

Страница 1: ... comp 10 lb tension 500 mV lb Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...ialized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For more information contact your local PCB Piezotronics distributor sales representative or factory customer service representative Returning Equipment If factory repair is required our representatives ...

Страница 3: ...ation on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precautions to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the operating instructions located in the manual This symbol indicates safety ea...

Страница 4: ...E 住房 O O O O O O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ... O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 6: ...stallation outline drawing and specification sheet at the back of this manual for details and dimensions of the particular sensor model number s purchased The following pages give a brief description of the various sensor series available recommended mounting procedures operation and recommended calibration In addition to PCB s line of ICP sensors each ICP sensor series outlined has corresponding ...

Страница 7: ...e ring mounting and preload requirements Figure 4 outlines some possible mounting configurations to which the ring series may be installed Figure 4 Series 201 to 207 ICP Ring Force Sensor Possible Installation Methods GENERAL PURPOSE RADIAL Model 208C01 C05 General Purpose Sensors are designed to measure compression and impact forces from a fraction of a lb N to 5 000 lbs to 22 24 kN Tension force...

Страница 8: ... simultaneously measuring force in three 3 orthogonal directions X Y and Z Internally three sets of quartz crystals are cut oriented and preloaded so the maximum output from the crystal cut for the specific plane is obtained when a force is applied axially to the specific x y or z plane Sensor output from loads applied in the alternate transverse planes considered channel cross talk is negligible ...

Страница 9: ...returning the sensor to PCB for recalibration Figure 8 SERIES 3 COMPONENT TRIAX FORCE SENSORS PENETRATION Penetration style sensors are similar to the axial models but are specifically designed for compression and impact force measurements in materials testing applications Smooth cylindrical housings and curved impact caps avoid cutting through specimens This enables measurements to be taken to de...

Страница 10: ...ot facing or surface grinding Surface flatness should be held to within 0 001 TIR over the entire mating surface When mounting sensors between two plates care should be taken to assure mounting surfaces are flat and parallel Non parallel surfaces could place unwanted stress on internal components leading to premature sensor failure The protective cap should remain on the electrical connector durin...

Страница 11: ...the left in Figure 13 part of the force between the two structures is shunted through the mounting stud The amount of force shunted is dependent on the stud material but may be up to 5 of the total force for the beryllium copper stud supplied with the sensor and up to 50 for steel studs This typical installation setup is used by PCB during standard calibrations of force ring models 201B01 through ...

Страница 12: ...ice the installer can monitor the sensor output voltage directly for correct preloading Monitor the output from the Z axis connector when preloading 3 component type sensors These sensors require a preload in the Z axis that is 10 times their shear range Some models require this to be accomplished in steps not to exceed the usable voltage To prevent clipping of the signal increments should not exc...

Страница 13: ...meters depending on the frequency of interest It is necessary to supply the sensor with a 2 to 20 mA constant current at 20 to 30 VDC through a current regulating diode or equivalent circuit contained in all PCB signal conditioners See Guide G 0001B for powering and signal conditioning information pertaining to all ICP instrumentation Most of the line powered signal conditioners manufactured by PC...

Страница 14: ...tant in the system For this reason the readout device as well as the signal conditioner must be DC coupled Figure 16 Characteristic Discharge Time Constant Curve The discharge time constant represents the decay rate of an input signal One DTC represents the amount of time taken for the signal to decay to 37 of the initial peak value As illustrated in Figure 16 this is an exponential decay Approxim...

Страница 15: ...turned to the factory for servicing in the event of serious malfunction Observe the following precautions in using the sensor A Do not exceed the maximum load levels for the force sensor see specification sheet B Do not subject the sensor to temperatures exceeding that of the specification normally 250 F 121 C C Do not apply voltage to the sensor without current limiting diodes or other current pr...

Страница 16: ...ck 10 32 Coaxial Jack Electrical Connection Position Axial Axial All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc N Negative Output Polarity Output Polarity Compression Negative Negative W Water Resistant Cable Electrical C...

Страница 17: ......