9
Individual periods can be enlarged by the integrated zoom function.
3
Charging the built-in accumulator
•
For charging the accumulator, use the provided original charger exclusively
•
Make sure that the charger‘s specified input voltage corrensponds to your local supply voltage.
•
The used power socket must be accessible throughout entire charging process.
• While charging, temperature must be b10°C and +45°C.
• The status LED‘s of the charger indicate charge state during process (red = charging, green = charging
finished)
•
Charging an empty accumulator takes about 3 hours.
•
You can inform yourself on accumulator‘s residual capacity by means of the PC software
PWB-Soft.
• To preserve the accumulator, you should deactivate measurement when not needed for some time by
setting measurement start ahead. The ideal storage residual capacity is about 20%, never store with an
empty accumulator.
Note: Depth discharge damages the accumulator permanently.
When falling below minimum voltage, data logging will be stopped. The data logger will be operated in power-
saving mode. If the accumulator‘s residual capacity falls below 20%, then, at the latest, it should be rechar-
ged. We recommend 100% charging prior to any operation.
The accumulator must not be replaced by user. If for your N6/60 a new accumulator is needed, forward the
device to the manufacturer for replacement.