Paul Wegener PWBlogg Скачать руководство пользователя страница 7

7

The setting „permanent limit value monitoring“ means permanent sampling in interval 2. This leads to an im-

mediate identification of threshold violations. Thereafter, the memory interval 1 will be redirected to interval 2. 

Please note that frequent sampling of transducers increases power consumption of data logger and decrea-

ses economic life-time of the battery!

Shorted alarm storage

In case of a threshold violation normally data recording will be continued with interval period 2. The period of 

data recording can be much shorter than expected. To avoid it, the option „shorted alarm storage“ can lower 

that period of time where interval 2 for storage is used. The storage is limited to a maximum of 2 x interval 

1 (eg 2 x 5 min). Subsequently, it is continued to measure in interval 2 but recorded just with the cycle of 

interval 1.

Password protection

The configuration can be protected via password against unauthorized access. Without knowledge of pass

-

word, it is not possible to clear stored data or to change configuration of the data acquisition system.

2.2 

Error sources

The analog signals of transducers are converted into a digital value with resolution of 8 bit, 12 bit, 14 bit or 

16 bit. The following table demonstrates the difference between resolutions by means of a 16 bar transducer. 

The values, represented in PWB-Soft program or on LCD must be rounded to the defaulted number of deci-

mal places so that the increment between two neighbouring values is not always identical.

Table: AD conversion, resolution

Potential error sources are non-linearities of AD converter which are irrelevant on resolution of 8 bit or 12 

bit. However, another important one is the digital step at small resolutions. At the limit of two sequent digital 

values, the tolerance between the measured (in digital steps presentable) and the real value (other sources 

of error excluded) already amounts ½ resolution step. As in the table 8 bit is equivalent to 31,25 mbar. Additio

-

nally, there are device-dependent sources of error such as temperature drop or deterioration of components. 

At resolutions as 14 bit and 16 bit, they have much importance.

 

 

 

 

 

 

Operating manual of data acquisition system PWBlogg

ADU

digital steps

reading resolution

8 bit

256

62,5 mbar

12 bit

4096

4 mbar

14 bit

16384

1 mbar

16 bit

65535

0,25 mbar

Содержание PWBlogg

Страница 1: ...Operating manual Data acquisition system PWBlogg Type N6 60 ...

Страница 2: ...Ballenstedt All rights reserved No section of the operating manual may be replicated or processed duplicated or distributed using electronic systems in any way print photocopy microfilm or other methods without prior written permission of the Paul Wegener GmbH Design and specifications are subject to change without notice 2 ...

Страница 3: ...unting and connecting the device must only be done if there is no potentially explosive atmosphere Connecting and disconnecting the plug connectors is only permitted if there is no explosive atmosphere Unused sockets must be closed with dedicated protective caps Unplugged connectors are not permitted in hazardous areas Always remove unplugged connectors from hazardous areas Opening the case must o...

Страница 4: ...er parameters 6 2 2 Error sources 7 2 3 Data communicationv 8 3 Charging the built in accumulator 9 4 Technical specifications 10 4 1 Data memory 10 4 2 Estimation of accumulator service life 11 5 Recycling batteries and old appliances 11 6 Troubleshooting 13 Paul Wegener GmbH 4 ...

Страница 5: ...r The data logger has an integrated analog digital converter with high resolution which is responsible for log ging and digitizing pressure and temperature readings A microprocessor stores the readings in the integra ted memory At its highest expansion stage it can log 3 8 million readings Data logger parameterizing as well as reading and evaluating performance data is done on any PC laptop or net...

Страница 6: ...erval 2 can be set from 1s optional 10ms up to 1min It will be activated once a threshold violation is recognized Threshold monitoring The data logger provides specifying thresholds and records threshold violations by means of a second sam ple rate Additional to the threshold a hysteresis can be specified Furthermore the data logger has an implemented sniff mode This mode is enabled in case alarm ...

Страница 7: ...ion system 2 2 Error sources The analog signals of transducers are converted into a digital value with resolution of 8 bit 12 bit 14 bit or 16 bit The following table demonstrates the difference between resolutions by means of a 16 bar transducer The values represented in PWB Soft program or on LCD must be rounded to the defaulted number of deci mal places so that the increment between two neighbo...

Страница 8: ...oped software PWB Soft provides data communucation with the data logger N6 60 Any necessary parameters on measurement are set here Fot that purpose this software provides different dialog windows Performance data evaluation can be done in tabular form or as chart ...

Страница 9: ...mulator s residual capacity by means of the PC software PWB Soft To preserve the accumulator you should deactivate measurement when not needed for some time by setting measurement start ahead The ideal storage residual capacity is about 20 never store with an empty accumulator Note Depth discharge damages the accumulator permanently When falling below minimum voltage data logging will be stopped T...

Страница 10: ...x 115200 Bd when reading data Power supply 7 2V lithium ion accumulator 3100 mAh 7 2V lithium battery 2100mAh optional Operating storage Temperature 20 60 C Explosion protection II 3G Ex ec ic Gc IIC T4 Gc X optional Environment 20 C Ta 60 C Electrical parameters for intrinsically safe status and counting inputs maximum values per circuit Voltage U0 9 2 V Current I0 9 3 mA Power P0 21 4 mW Capacit...

Страница 11: ... are subject to particular statutory provisions Batteries rechargeable batteries and electric appliances must not be disposed with the household waste The end user is obligated to return them by the law Used electric appliances batteries and rechargeable batteries can be disposed at collecting points municipal disposal areas or by manufacturer supplier The Paul Wegener GmbH as seller of batteries ...

Страница 12: ...ference resistance industrial sector EN 60079 0 2019 09 electrical equipment for potentially explosive atmospheres section 0 general requirements EN 60079 7 2016 08 potentially explosive atmospheres section 7 equipment protection by increased safety e EN 60079 11 2012 06 explosive atmosphere section 11 equipment protection by intrinsic safety i Marking as category 3 equipment for use in gas explos...

Страница 13: ...een intruded into the data logger Accumulator is low Have it replaced by manufacturer Password forgotten Read out performance data send reading file to manufacturer Software does not read performance data comple tely or reports errors when reading out the data Check whether the PC data cable is properly connected Have checked by manufacturer whether moisture is intruded in the data logger accumula...

Страница 14: ...Paul Wegener GmbH 14 ...

Страница 15: ...15 Operating manual of data acquisition system PWBlogg ...

Страница 16: ...GmbH 16 Paul Wegener GmbH Marienstraße 24 D 06493 Ballenstedt Tel Fax Internet e mail 49 0 39483 96 300 49 0 39483 96 400 www paul wegener de info paul wegener de PWB FORM 2084 EN22 Most recent emendment 16 06 2022 ...

Отзывы: