Wireless Drop Counter

Assembling the Titration Setup (See figure on next 

12

013-16026A

15.

Touch or click “New 
Calibration” to see the Slope 
and Offset values. In this 
example, the Slope is 
“2.0000 drops/mL” because 
10 drops were counted and 
the volume of the titrant you 
entered is 5 mL.

16.

Click or touch “OK” to close 
the “Calibrate Sensor” 
window. Click or touch in the 
Graph display to close the 
“Configure sensors and 
measurements” menu.

About the Calibration

Calibration information is stored 
in the Wireless Drop Counter 
Sensor and can be used by both 
PASCO Capstone and 
SPARKvue. The information can 
be changed as needed. Note 
that if you use a different liquid 
as the titrant, or if you change 
the drop rate or drop size by 
adjusting the stopcock valves on 
the Drop Dispenser, you will need to re-calibrate.

Acid-Base Titration (Calibration Method)

1.

Setup the equipment and consumables as shown. 

2.

Start the data acquisition software.

3.

In the software, “pair” (connect) the Wireless Drop Counter and a Wireless pH Sensor

4.

Use the software to create a graph display of pH (vertical axis) vs. Fluid Volume (horizontal axis).

5.

Start the magnetic stirrer. Check that the micro stir bar is able to rotate.

6.

In the software, touch, or click “Start” or “Record” to begin recording data.

7.

Adjust the stopcock valves on the Drop Dispenser so 
the titrant drops fall at about 1 drop per second.

8.

Observe the data on the pH versus Fluid Volume graph. 
After the equivalence point is reached, continue 
collecting data until the pH curve flattens.

9.

Close the stopcock valve and stop data recording. 

In the software, touch or click the Start (or Record) 
button to begin collecting data.

 Touch or Click to see the results

The “Slope:” is 2.0000 because the sensor 

counted 10 drops and you entered 5.0 mL 

as the volume

Equivalence point

Содержание PS-3214

Страница 1: ...Reference Guide 013 16026A Wireless Drop Counter PS 3214 a b d c e f Power Rechargeable Battery or USB Cable Connection...

Страница 2: ...t or smart phone when the sensor is connected PASCO Capstone will work on a PC or Mac The Wireless Drop Counter can be mounted on a support rod The counter is equipped with two integrated probe holder...

Страница 3: ...blocking the drop window To turn the sensor on press and hold the ON button When the battery status LED shines red for a moment release the ON button Then the Bluetooth status LED will blink This ind...

Страница 4: ...elp from the Help menu or press F1 LED Information The drop LED operates as follows The drop LED blinks when a drop is detected passing through the drop window If the lens in the drop window becomes o...

Страница 5: ...in the liquid in the beaker Assemble the Drop Dispenser 1 Remove the plunger from the syringe Remove the protective white plastic caps from the stopcock valves 2 Connect the two stopcocks together Con...

Страница 6: ...e figure on next 6 013 16026A Syringe with 50 mL of 0 1 M NaOH Wireless pH Sensor Direct Connect pH Probe Wireless Drop Counter Drop Tip Stopcock Stopcock Syringe Holder Support Rod Micro Stir Bar Mag...

Страница 7: ...drops from the drop tip and give you a measure of the volume 5 Start Capstone Pair the Wireless Drop Counter wirelessly to the computing device e g computer In Capstone open the Hardware Setup window...

Страница 8: ...om stopcock so that the drops fall at a rate of about 1 drop per second Close the top valve after the rate is achieved Discard the titrant as directed by your instructor 4 Align the drop tip with the...

Страница 9: ...me is the measurement selected as default The Templates window also opens with Graph Table and Graph and Digits as the choices 8 In the Templates window touch or click the Graph template to open a Gra...

Страница 10: ...e that the Fluid Volume measurement also has a Calibrate measurement icon 10 Click or touch the Calibrate measurement icon to open the Calibrate Sensor Select Measurement window 11 In the Calibrate Se...

Страница 11: ...the number of drops as they are counted 13 After about 5 mL of titrant has collected in the graduated cylinder close the bottom stopcock Carefully measure the exact volume of titrant in the graduated...

Страница 12: ...ibrate Acid Base Titration Calibration Method 1 Setup the equipment and consumables as shown 2 Start the data acquisition software 3 In the software pair connect the Wireless Drop Counter and a Wirele...

Страница 13: ...d then back on Retry If the Wireless Drop Counter will not turn on use the micro USB cable to connect it to a USB port or charger See the PASCO web site at www pasco com for more information Probe Car...

Страница 14: ...e that it will be recycled in a manner that protects human health and the environment To find out where you can drop off your waste equipment for recycling please contact your local waste recycle disp...

Страница 15: ...he System Report button Select Bluetooth from the sidebar on the left underneath Hardware Scan down the list of information until you find LMP Version If your Mac is equipped with Bluetooth SMART the...

Страница 16: ...Wireless Drop Counter Appendix A Bluetooth Compatibility 16 013 16026A...

Отзывы: