deionized water and gently wipe the interior. 

The cuvettes included with this spectrometer 

are made from polystyrene. With these 

cuvettes, only use solutions to which 

polystyrene shows good chemical resistance. 

Battery: 

Life

On a full charge, the Spectrometer will run in 

Analyze Light mode for eight to nine hours. 

The battery is replaceable. 

Replacement 

If the Spectrometer fails to recharge, the 

battery may need to be replaced. The battery 

is field replaceable using the Model PS-2569 

Replacement Battery. Removing the battery 

requires a Torx T8 screwdriver and a small 

Phillips head screwdriver.
CAUTION: Make the area as dust-free and 

static-free as possible before opening the 

Sprectrometer. Do not touch the optical 

components inside the Spectrometer.
• Use the Torx screwdriver to remove the four 

screws from the bottom of the 

Sprectrometer case.

• Lift the top half of the case off the 

Spectrometer and set it aside.

• Disconnect the battery leads from the 

connector on the bottom edge of the printed 

circuit board.

• Use a small Phillips head screwdriver to 

remove the one small screw that is near the 

middle of the top edge of the printed circuit 

board.

• Lift the printed circuit board aside, and 

replace the square shaped battery pack. 

Reconnect the battery leads to the 

connector on the bottom edge of the printed 

circuit board.

• Check that the printed circuit board and its 

components are free of dust.

• Reverse the other steps to close the 

Spectrometer.

Contact PASCO Technical Support for more 

information.

Light Source Replacement

The small printed circuit board that holds the 

incandescent bulb and the three light-emitting 

diodes can be removed and replaced by a 

skilled technician. Contact Technical Support 

regarding availability and installation 

instructions for PASCO part number 

004-14026. 

Troubleshooting the Spectrometer

• If the Spectrometer loses Bluetooth 

connection and will not reconnect, try 

cycling the ON/OFF button. Press and 

briefly

 hold the button until the status LEDs 

blink in sequence, and then release the 

ON/OFF button to turn on the 

Spectrometer. 

• If the Spectrometer stops communicating 

with the computer software or tablet 

application, try restarting the software or 

application. If the problem remains, try 

cycling the ON/OFF button. 

• As a last resort, push a pin into the Reset 

port on the left end of the Spectrometer. 

Technical Support

For assistance with any PASCO product, 

contact PASCO at:

For more information about the product and 

the latest revision of this Reference Guide, 

visit the PASCO web site at: 

www.pasco.com/manuals
and enter “Spectrometer” or the product 

number (PS-2600) in the Search window.

Limited Warranty

The Spectrometer has a five year warranty. For a more thorough 

description of the product warranty, see the PASCO catalog.