background image

L55 and IL55BV2 Sensor Instruction Manual

12  Appendix 1: Preparation of Formazin Turbidity Standard

12.1  Health & Safety Precautions

The chemicals used when following this procedure are harmful, therefore the correct safety precautions must be 
carried out. During handling, avoid inhalation and contact with the eyes or skin. Wash hands thoroughly after use. 
Refer to the data sheets for further details of the chemicals used.

12.2  Method of Preparation

The preparation of a Formazin Turbidity Standard should be carried out with great care. Absolute cleanliness should 
be observed at all times to avoid any contamination of the solution.

All water used should be freshly distilled or de-mineralised. During preparation and use, keep the solution covered to 
prevent contamination.

1. Dissolve the Hydrazinium Sulphate 10g (32112680) into 1 litre of water. Stir the solution thoroughly.

2. Dissolve the Hexamine 100g (32112630) into 1 litre of water. Stir the solution thoroughly.

3. Pour the above solutions into a 2.5 litre Winchester or similar type amber glass bottle and mix thoroughly. Allow the 

mixture to stand undisturbed for 24 hours at 25

°

C. The resultant solution will have a turbidity of 4000 FTU.

4. Prior to use the standard should be mixed thoroughly, i.e. before pouring out the standard, shake the bottle well.

12.3  Preparation of Dilution’s.

The 4000 FTU standard can be used to prepare solutions of lower values by diluting with water. The formula below 
can be used to calculate the ratio of water to 4000 FTU standard to produce a solution of the required value.

W

A

=

4000

1

W = parts water to 1 part 4000 FTU standard

A = Required FTU value 

The Formazin Dilution Chart (table QAP 6.14.4) provides the mixture requirements to make various solutions using the 
4000 FTU standard.

12.4  Calibration

When Formazin solutions are used for calibration/testing of sensors, sufficient solution is required such that the sensor 
is completely immersed. When using light scatter sensors it is important that the sensor is positioned away from the 
sides and bottom of the container to prevent the light being reflected. 

12.5  Storage Life

The turbidity standard should be stored in a Winchester bottle and kept in a cool dark place. The bottle should be 
labelled to identify its content and also the date of preparation should be recorded. The 4000 FTU standard has a shelf 
life of one year after which it should be discarded and a fresh solution should be prepared.

Solutions made by diluting the 4000 FTU standard should also be stored in suitably labelled amber glass bottles and 
kept in a cool dark place. Solutions above 400 FTU have a shelf life of one month after which they should be 
discarded. Solutions below 400 FTU should be prepared from the 4000 FTU standard daily.

Solutions can be used outside of the storage lives specified above but any results taken must acknowledge this and 
be accepted as less accurate. When using solutions, watch for flocculation (the particles in the solution will appear to 
link together), if this occurs discard the solution and prepare a fresh standard.

163790IM-Iss05 Issue Date 11/05/210

Page 19 of 22

Содержание IL55

Страница 1: ...INSTRUCTION MANUAL IL55 and IL55BV2 Sensor In Line Sludge Density and Turbidity Sensor...

Страница 2: ...This page is intentionally left blank...

Страница 3: ...bration 13 7 Maintenance 14 7 1 Sensor Cleaning 14 7 2 Inspection 14 7 3 Fault Finding 14 8 Application Example Sludge Density Monitor 15 9 Spare Parts 16 9 1 Sensors 16 9 2 Mounting Accessories 16 9...

Страница 4: ...d to ensure long term reliability The sensor uses 180 Back Scatter of infrared light to detect the presence of solids in the pipe the sensor electronics provides an output signal that is used by eithe...

Страница 5: ...be completely flooded at all times There should be no rapid changes in pipe size near to the sensor installation point Do not place the sensor in an area of no flow as the suspended solids will settl...

Страница 6: ...existing pipe section Depending upon pipe material the boss will be welded steel or adhered plastic to the parent material A hole to suit the inside diameter of the boss will be required through the...

Страница 7: ...problems with the sensor measurement Good engineering practice should be followed with particular attention being paid to the following points The sensor cable should be kept to the minimum length po...

Страница 8: ...ns that have a maximum line pressure of 2 Bar or less Page 8 of 22 163790IM Iss05 Issue Date 11 05 2010 Mounting Boss Ball Valve Sensor shaft Ball valve coupling nut Dowty Seal Dowty seal retaining nu...

Страница 9: ...n positioned and fastened to the pipeline the ball valve mechanism via coupling nut can be screwed into the mounting boss 2 Insert the sensor shaft into the ball valve mechanism by screwing and tighte...

Страница 10: ...process as outlined in the previous section 1 Loosen the shaft locking nut 2 Ensure that there is sufficient free cable to allow sensor removal 3 Withdraw sensor to end stop 4 Close ball valve 5 Unscr...

Страница 11: ...G It must be noted that the IL55 can only be used in applications that have a maximum line pressure of 5 Bar or less 163790IM Iss05 Issue Date 11 05 210 Page 11 of 22 Mounting Boss Threaded Collar Fit...

Страница 12: ...carrying out the procedure below 1 Loosen the knurled ring 2 Ensure that there is sufficient free cable to allow the sensor to be removed 3 Undo the knurled ring do rotate the sensor whilst doing thi...

Страница 13: ...the monitor instruction manual the following precautions must be carried out with the IL55 IL55BV2 sensors to ensure accurate calibration The sensor must be held in a dark light proof container for z...

Страница 14: ...dried onto the optical surface a detergent or weak acid solution can be used Care must be taken not to damage the optical surface 7 2 Inspection It is envisaged that an inspection of the sensor will i...

Страница 15: ...information that allows the user to control the de sludging of the tank The instrument will warn of a low level of suspended solids being pumped from the tank allowing the user to stop pumping and al...

Страница 16: ...520 IL55BV2 Sensor 9 2 Mounting Accessories 171310 IL55 Mounting Boss Stainless Steel 171300 IL55BV2 Mounting Boss Stainless Steel For pipe sections please consult Partech Instruments 9 3 Miscellaneou...

Страница 17: ...Description of fault 10 1 Returning Equipment for Repair If equipment needs to be returned to Partech for repair or service the following address should be used SERVICE DEPARTMENT PARTECH ELECTRONICS...

Страница 18: ...Polyurethane Glass Weight IL55 2 5 Kg IL55BV2 7 2 Kg Dimensions IL55 Length 190mm IL55BV2 Length 392mm 695mm extended Standard Cable Length 10 metres Maximum Cable Length 100 metres EMC Emmision EN 50...

Страница 19: ...ed to calculate the ratio of water to 4000 FTU standard to produce a solution of the required value W A 4000 1 W parts water to 1 part 4000 FTU standard A Required FTU value The Formazin Dilution Char...

Страница 20: ...ion Required g l Add pack to water litres 2 5 8 0 5 0 4 0 10 0 2 0 20 0 1 0 40 0 0 5 The following points should be observed when preparing and using Fullers Earth solutions The water used should be f...

Страница 21: ...L55 and IL55BV2 Sensor Instruction Manual 163790IM Iss05 Issue Date 11 05 210 Page 21 of 22...

Страница 22: ......

Отзывы: