6  Maintenance

To maximise the operational life of the monitor and sensor, regular inspection is recommended. The frequency of this 
is dependent on the specific operation of each site, however an initial weekly inspection to observe the characteristics 
of the application is recommended, long term operational practice can be decided from these inspections.

The manual clean system requires the physical removal of the sensor from the application for the cleaning of the 
lenses to be carried out. 

It is advisable that at regular intervals the sensor is removed from the application and both the lenses and the sensor 
body are cleaned along with the mounting brackets.

6.1  Inspection

It is envisaged that an inspection of the monitor and sensor will include the following:

Check security of brackets

Removal of the sensor from the application.

Checking for damage to cables.

Cleaning of the sensor using a suitable detergent and removal of all ragging.

Inspection of the glass lenses for staining, scouring, cracks or chips.

Checking for moisture/signs of leakage inside the sensor and monitor.

6.2  Fault Finding

The following can cause a false alarm or fault signal:

Material (rag, weed, etc) caught on the sensor and affecting the measuring system - remove sensor, manually 
clean and re-install.

Turbulence   (caused   by  high   flow,   aeration   etc)  in   the  solution   being  measured   –   position   the   sensor  so 
turbulence has as minimal effect as possible on the sensor.

Bad connection of the sensor cable - check all connections are securely fastened

168270IM-02 Issue Date 05/01/2010

Page 11 of 14

Содержание 8100

Страница 1: ...INSTRUCTION MANUAL 8100 8200 Monitor Version 3...

Страница 2: ...This page is intentionally left blank...

Страница 3: ...ical Connections 7 3 1 Supply Voltage 7 4 Configuration 8 4 1 Setpoint 8 4 2 Delay 8 5 Sludge Blanket Level Detection 9 5 1 Description 9 5 2 Sensor Installation 9 5 3 Monitor Configuration 9 6 Mainte...

Страница 4: ...alarm set point and delay for each channel can be independently set and the status of each channel is indicated separately The 8100 8200 Monitor has an IP65 rated enclosure and is suitable for outdoo...

Страница 5: ...0 Monitors are therefore suitable for interfacing with automatic de sludging control of primary and secondary settlement tanks thickener control and high and low blanket level alarm or as a simple ala...

Страница 6: ...0 Monitor is designed so that it can be surface mounted the hole positions required for this are shown below Mounting hardware such as handrail brackets protective hoods and sensor brackets are availa...

Страница 7: ...The instrument has a supply fuse fitted internally The 8100 monitor is supplied with three PG11 cable glands fitted and has provision for a further PG11 gland whereas the 8200 comes with four fitted c...

Страница 8: ...the system once installed By turning the setpoint control from fully anti clockwise to fully clockwise it is possible to alter the setpoint from 5 95 of the range of the sensor 4 2 Delay The delay con...

Страница 9: ...ining clamps and lowering it into the settlement tank to the desired depth at which the sludge blanket is required to be detected When assessing mounting options attention should be paid to the access...

Страница 10: ...es the presence of the sludge blanket when the sensor is above the blanket the sensor that is fitted is too sensitive and an alternative sensor will have to be fitted For accurate adjustment of the se...

Страница 11: ...6 1 Inspection It is envisaged that an inspection of the monitor and sensor will include the following Check security of brackets Removal of the sensor from the application Checking for damage to cab...

Страница 12: ...Application details Description of fault 7 1 Returning Instruments for Repair If equipment needs to be returned to Partech for repair or service the following address should be used SERVICE DEPT PART...

Страница 13: ...ge 0 to 50 C Storage Temperature Range 20 to 60 C 8 1 2 Alarm Outputs Number of Outputs 8100 1 alarm 1 fault 8200 2 alarms 1 fault Alarm Type SPCO Operation Fail Safe De energised in alarm condition R...

Страница 14: ......

Отзывы: