8100/8200 Monitor Instruction Manual

5.3.2 Blanket continually detected

If the sensor is continually detecting the sludge interface and the red ALARM LED stays on when the sensor is above 
the sludge blanket the following action should be taken:

With the sensor above the sludge blanket, set the delay to the minimum, then the alarm point control should be 
slowly   turned   clockwise   until   the   green   OK   LED   illuminates.   This   indicates   that   the   sensor   is   correctly,   not 
detecting the sludge blanket.

Note: If the setpoint control is turned fully clockwise and the detector continually indicates the presence of the sludge 
blanket when the sensor is above the blanket, the sensor that is fitted is too sensitive and an alternative sensor will 
have to be fitted
For accurate adjustment of the setpoint to a required level:

1. Immerse the sensor fully in a known sample of the required value.

2. Turn the alarm point control fully anti-clockwise; the green OK LED should illuminate.

3. The alarm point control should be turned clockwise until the red ALARM LED illuminates.

The desired alarm point has now been set and the controller will now detect sludge blankets of this value or higher.

Note: In the event of the sensor being changed, it will be necessary to repeat the above alarm point adjustment in 
order to obtain accurate results

168270IM-02 Issue Date 05/01/2010

Page 10 of 14

Содержание 8100

Страница 1: ...INSTRUCTION MANUAL 8100 8200 Monitor Version 3...

Страница 2: ...This page is intentionally left blank...

Страница 3: ...ical Connections 7 3 1 Supply Voltage 7 4 Configuration 8 4 1 Setpoint 8 4 2 Delay 8 5 Sludge Blanket Level Detection 9 5 1 Description 9 5 2 Sensor Installation 9 5 3 Monitor Configuration 9 6 Mainte...

Страница 4: ...alarm set point and delay for each channel can be independently set and the status of each channel is indicated separately The 8100 8200 Monitor has an IP65 rated enclosure and is suitable for outdoo...

Страница 5: ...0 Monitors are therefore suitable for interfacing with automatic de sludging control of primary and secondary settlement tanks thickener control and high and low blanket level alarm or as a simple ala...

Страница 6: ...0 Monitor is designed so that it can be surface mounted the hole positions required for this are shown below Mounting hardware such as handrail brackets protective hoods and sensor brackets are availa...

Страница 7: ...The instrument has a supply fuse fitted internally The 8100 monitor is supplied with three PG11 cable glands fitted and has provision for a further PG11 gland whereas the 8200 comes with four fitted c...

Страница 8: ...the system once installed By turning the setpoint control from fully anti clockwise to fully clockwise it is possible to alter the setpoint from 5 95 of the range of the sensor 4 2 Delay The delay con...

Страница 9: ...ining clamps and lowering it into the settlement tank to the desired depth at which the sludge blanket is required to be detected When assessing mounting options attention should be paid to the access...

Страница 10: ...es the presence of the sludge blanket when the sensor is above the blanket the sensor that is fitted is too sensitive and an alternative sensor will have to be fitted For accurate adjustment of the se...

Страница 11: ...6 1 Inspection It is envisaged that an inspection of the monitor and sensor will include the following Check security of brackets Removal of the sensor from the application Checking for damage to cab...

Страница 12: ...Application details Description of fault 7 1 Returning Instruments for Repair If equipment needs to be returned to Partech for repair or service the following address should be used SERVICE DEPT PART...

Страница 13: ...ge 0 to 50 C Storage Temperature Range 20 to 60 C 8 1 2 Alarm Outputs Number of Outputs 8100 1 alarm 1 fault 8200 2 alarms 1 fault Alarm Type SPCO Operation Fail Safe De energised in alarm condition R...

Страница 14: ......

Отзывы: