background image

Operating Basics 

 

Rev. 007 Date 2010-01-13

 

   

CyFlow

®

 space Instrument Operating Manual 

9/26

 

True Volumetric Absolute Counting 

– Overview 

 
 
Analysis  of  concentrations  of  samples,  loosely 
called  Absolute  Counting,  can  be  of  significant 
interest  for  medical  diagnostics  or monitioring  the 
status  of  cells  in  cell  cultures  or  biotechnological 
processes.  
 

Traditional  Flow  Cytometric  Techniques  for 
Absolute Counting 

 
i) Dual Platform Technique 
 
To  determine  the  concentration  of  a  sample  flow 
cytometers  (FCM)  are  frequently  used  in 
combination with Hematology Counters.  
 
Disadvantages: 

 

Two instruments required 

 

This  technique  relies  on  the  assumption  that 
the  number  of  cells  seen  by  the  FCM  and 
counted  by  the  Hematology  Counter  are 
identical. 

 
ii)  Single  Platform  Technique  with  Reference 
Beads 
 
An  alternative  method  employes  reference  beads 
of  known  concentration  which  are  added  to  the 
sample of unknown concentration. 
 
Disadvantages: 

 

Depends  on  the  accuracy  of  the  specified 
beads concentration 

 

Constant  running  costs  for  the  reference 
beads  

 

Partec True Volumetric Absolute Counting 

 
To  overcome  the  drawbacks  of  the  Dual-  and 
Single-Platform  Technique  with  reference  beads, 
Partec  intruments  additionally  offer  an  alternative 
way  of  absolute  cell  counting,  which  is  based 
directly on the basic definition of a concentration 
  
 
by  precisely  counting  the  number 

N

  of  cells 

suspended  in  a  purely  mechanically  defined 
volume 

V

.  

 
i) Precise Counting: Determination of 

N

 

 
For  True  Volumetric  Counting  the  precise 
detection of cells is an essential. This requires fast 
recognition  and  analysis  of  the  events  by 
electronics  and  computer.  All  Partec  instruments 
are  specifically  designed  to  minimize  counting 
losses  by  providing  direct  connection  between 

computer  and  electronics,  which  avoids  dead-
times  involved  in  traditional  FCM  designs  and 
instrument interfaces. This reduces the probability 
of  a  counting  loss  for  typical  event  rates  below 
2%. 
 
ii) Electrode-Principle: Determination of 

V

 

The  method  for  True  Volumetric  Absolute 
Counting  supported  by  the  CyFlow

®

  space  is 

based  on  the  precise  measurement  of  a  fixed 
sample volume by means of two electrodes.  
 

 

Fig.  1:  True  Volumetric  Absolute  Counting  by  the 
Electrode Principle. 

 
During  analysis  by  the  CyFlow

®

  space,  the 

sample  liquid  loses  contact  first  from  the  upper 
and  then  from  the  lower  electrode.  These  events 
trigger a  START and a STOP signal, respectively 
which is stored by software. 
The volume 

V

 of sample liquid analysed between 

the START and STOP signal is physically defined 
by  the  distance  of  the  two  electrodes  and  the 
diameter  of  the  sample  tube.  The  analysed 
volume is exactly 200 µl (= counting volume). 
 
Effects  of  the  sample  meniscus  in  the  tube  are 
eliminated  because  the  START  and  STOP 
electrodes  are  arranged  symmetrically.  The 
sample  conductivity  does  not  influence  the 
volumetric measurement, as long as the fluids can 
be detected electrically.  
 
In some cases, larger or smaller counting volumes 

V

  are  advantageous,  either  to  improve  the 

counting statistics (e.g. for rare events < 10/ml) or 
to  decrease  the  analysis  time.  This  can  be 
accomplished  by  replacing  the  standard  200 µl 
volumetric  sample  port  by  one  with  a  different 
volume  (Please  contact  Partec  or  your  local 
distributor for details). 

 

c

 = 

N

 / 

V

 

,

 

Содержание CyFlow space

Страница 1: ...CyFlow space Instrument Operating Manual...

Страница 2: ...ent Settings 12 The Parameter Setup Dialog Box Pulse Height Area and Width 13 Trigger 14 PMT High Voltage Gain and Log Amplification 15 Sample Speed 16 Threshold Lower Level L L 17 Appendix 18 Install...

Страница 3: ...lly any flow cytometric application The applications cover e g Routine and research Immunophenotyping Blood Cell Analysis HIV monitoring Leukocyte Counting Rare Event Analysis Microorganism Analysis F...

Страница 4: ...haracteristic colour emission wavelength spectrum This fluorescence light is separated into colour ranges by means of optical filters The intensity of each colour range is analysed for each single cel...

Страница 5: ...ometer is operated with 100 240 V AC Switch on main power at the left side of the instrument The 488nm laser has its own power button closeby please also switch on the 488 nm laser All other lasers ar...

Страница 6: ...s of the flow cuvette The CyFlow space Flow Cytometer supports either 2 laser spots when employing FloMax version 2 6 or 2 7 or 3 laser spots when employing FloMax version 3 0 The blue 488 nm laser wh...

Страница 7: ...nsert sample tube onto the sample port until you recognize a click The sample should be fully mounted within a second Now the measurement acquisition starts automatically the operating software indica...

Страница 8: ...ath fluid tube for 5 seconds Stop the instrument while keeping the tube pinched Incubate for 15 minutes Re start the system by pressing START and let it run to the end Run the system with 1 6 ml Parte...

Страница 9: ...ential This requires fast recognition and analysis of the events by electronics and computer All Partec instruments are specifically designed to minimize counting losses by providing direct connection...

Страница 10: ...No additional analysis steps e g setting gates for beads are required v Less Expenses No reference beads required Performing True Volumetric Absolute Counting Sedimentation and Count Time Cells or oth...

Страница 11: ...rward scatter SSC side scatter FL1 green fluorescence FITC FL2 orange fluorescence PE FL3 red fluorescence I PE Cy5 FL4 far red fluorescence I PE Cy7 FL5 red fluorescence II APC FL6 far red fluorescen...

Страница 12: ...ion for the particles of interest The adjustments cover the gains of the optical detectors e g the photomultiplier high voltages the amplification mode lin 3 or 4 decade logarithmic lower and upper le...

Страница 13: ...meter label To change parameter names click into the fields and enter the names by using the keyboard For each optical channel you may analyse three different pulse properties 1 The pulse height 2 the...

Страница 14: ...l other particles will not be recognized by the instrument The trigger parameter can be used for an efficient exclusion of unwanted particles from the analysis Example Assume vertebrate leukocytes in...

Страница 15: ...the instrument settings box see Fig 4 By using the Right Left buttons gain values are increased decreased In case linear amplification is used peaks are expanded to the right compressed to the left w...

Страница 16: ...decreased by clicking into the corresponding field in the instrument settings box and using the Right Left buttons see Fig 4 The count rate increases with elevated speed values If the speed is increas...

Страница 17: ...of small and unwanted background or noise signals below a threshold The L L value of a parameter can be increased decreased by clicking into the corresponding L L value field in the instrument setting...

Страница 18: ...educe smoke dust vibrations direct sunlight and direct neighbourhood of heatings as possible The installation room must be well ventilated and dry Temperature 15 30 C Humidity 20 85 relative non conde...

Страница 19: ...space shown from the rearside CyFlow rearside Screen Computer rearside Printer CCD Camera Interface Monitor Printer CCD Computer AC AC AC Air Sheath Mouse Keyboard Sheath Waste AC Waste CyFlow rearsid...

Страница 20: ...e supply sheath pressure and sample flow rate The Computer connections of the CyFlow space are located at the rear side of the instrument Connect the CyFlow space to the computer with the interface ca...

Страница 21: ...during sample run remove your sample and activate CLEAN several times In this way the flow cuvette is back flushed with sheath fluid Restart the sample and observe peak resolution again If the results...

Страница 22: ...lled water and a clean brush and flush with clean destilled water several times Remember cleanliness of sheath fluid reservoir is critical for proper operation If the CyFlow space will not be used for...

Страница 23: ...1 2001 Warning Laser light can be emitted if the protection cover for the laser beam is removed and the beam shutter is opened Therefore the system is marked with the following laser safety labels War...

Страница 24: ...entation control Particle Concentration Analysis True Volumetric Absolute Counting Particle Size and Fluorescence Distribution Analysis True Volumetric Based on precise counting and mechanical fluid v...

Страница 25: ...d precision syringe pump for contamination free sample transport Built in air pressure for sheath fluid Sheath fluid pressure is adjustable from 0 300 mbar Computer controlled Default setting 200 mBar...

Страница 26: ...d to 32 gates in free logical combinations Regions quadrants 1P ranges Gating Crosstalk On or offline gating and crosstalk compensation provide adjustments without need Compensation to rerun samples D...

Отзывы: