background image

 

Rev. 007 Date 2010-01-13

 

   

CyFlow

®

 space Instrument Operating Manual 

4/26

 

 Typical 

Steps 

of 

Particle 

Analysis    

 

 

 

 
 

1.  Cells  are  separated  in  a  suspension  and 
stained with fluorescent markers. 

 
 
 
 
 
 

 

 
 

2.  Light  from  the  laser(s)  excites  cell 

–bound 

fluorophores and is scattered by the cells. 

 
 
 
 
 

 

 

3.  Fluorescence  signals  are  displayed  and 
analysed 

in 

histogram- 

and/or 

dotplot-

diagrams. 
 
 
 

1. Preparation and Staining 

For  flow  cytometric  analysis,  the  cells  (or  other 
particles)  must  be  separated  from  each  other  in 
an  aqueous  suspension.  Typically,  if  not 
generating 

sufficient 

optical 

signals 

by 

themselves, the cells are then labelled by staining 
with  one  or  more  fluorescent  dyes.  The 
fluorescent  molecules  bind  to  the  cell  substance 
of  interest,  e.g.  CD4-  and  CD45-antibodies  bind 
cell  surface  antigens  of  leukocytes.  The  quantity 
of labelling cells with a specific dye is proportional 
to  the  number  of  antigen  molecules  of  a  single 
cell.  A  good  preparation  with  suited  reagents  is 
the requirement for a precise analysis. A check of 
staining  in  a  fluorescence  microscope  can  un-
reveal preparation problems. 
 

2. Flow Cytometry Analysis 

While passing through a flow cuvette one-by-one, 
the  cells  are  individually  illuminated  by  the  light 
spot  of  the  laser.  Due  to  the  excitation,  the  dye 
molecules  emit  fluorescence  of  characteristic 
colour  (emission  wavelength  spectrum).  This 
fluorescence  light  is  separated  into  colour  ranges 
by  means  of  optical  filters.  The  intensity  of  each 
colour range is analysed for each single cell.  
Besides  fluorescence,  the  intensity  of  light 
scattered  by  each  cell  can  be  measured.  Scatter 
light  is  measured  in  forward  direction  from  the 
light  source  (forward  scatter,  FSC)  and  sideward 
direction (side scatter, SSC). The scatter intensity 
is a measure of cell size and morphology. Scatter 
light  can  be  used  to  identify  a  cell  before 
analysing  its  fluorescence,  but  a  cell  can  also  be 
identified  by  fluorescence  before  analysing  its 
scatter properties. 
 

3. Realtime Data Processing and Results 

The  light  intensity  of  each  parameter,  and  with  it 
the  quantity  of  substance  of  interest,  is  assigned 
to  one  of  up  to  2

16

  or  65536  quantity  classes 

(channels).  The  classification  is  performed  in 
realtime,  while  the  cells  are  passing  the  flow 
cuvette.  Single  parameter  histograms  show  the 
number  of  cells  within  the  channels.  Two 
parameter  dotplots  show  the  correlation  between 
two cell properties.  
 

4. Absolute Cell Counting 

Since  the  CyFlow

®

  space  analyses  all  cells 

passing  through  the  flow  cell  while  precisely 
monitoring  the  fluid  volume  of  the  sample,  it 
allows volumetric counting during the analysis, i.e. 
the  determination  of  concentration  of  any  cell 
subpopulation.  Cell  subpopulations  can  even  be 
defined and their concentration analysed at a later 
date, after reloading data from a file. 

Содержание CyFlow space

Страница 1: ...CyFlow space Instrument Operating Manual...

Страница 2: ...ent Settings 12 The Parameter Setup Dialog Box Pulse Height Area and Width 13 Trigger 14 PMT High Voltage Gain and Log Amplification 15 Sample Speed 16 Threshold Lower Level L L 17 Appendix 18 Install...

Страница 3: ...lly any flow cytometric application The applications cover e g Routine and research Immunophenotyping Blood Cell Analysis HIV monitoring Leukocyte Counting Rare Event Analysis Microorganism Analysis F...

Страница 4: ...haracteristic colour emission wavelength spectrum This fluorescence light is separated into colour ranges by means of optical filters The intensity of each colour range is analysed for each single cel...

Страница 5: ...ometer is operated with 100 240 V AC Switch on main power at the left side of the instrument The 488nm laser has its own power button closeby please also switch on the 488 nm laser All other lasers ar...

Страница 6: ...s of the flow cuvette The CyFlow space Flow Cytometer supports either 2 laser spots when employing FloMax version 2 6 or 2 7 or 3 laser spots when employing FloMax version 3 0 The blue 488 nm laser wh...

Страница 7: ...nsert sample tube onto the sample port until you recognize a click The sample should be fully mounted within a second Now the measurement acquisition starts automatically the operating software indica...

Страница 8: ...ath fluid tube for 5 seconds Stop the instrument while keeping the tube pinched Incubate for 15 minutes Re start the system by pressing START and let it run to the end Run the system with 1 6 ml Parte...

Страница 9: ...ential This requires fast recognition and analysis of the events by electronics and computer All Partec instruments are specifically designed to minimize counting losses by providing direct connection...

Страница 10: ...No additional analysis steps e g setting gates for beads are required v Less Expenses No reference beads required Performing True Volumetric Absolute Counting Sedimentation and Count Time Cells or oth...

Страница 11: ...rward scatter SSC side scatter FL1 green fluorescence FITC FL2 orange fluorescence PE FL3 red fluorescence I PE Cy5 FL4 far red fluorescence I PE Cy7 FL5 red fluorescence II APC FL6 far red fluorescen...

Страница 12: ...ion for the particles of interest The adjustments cover the gains of the optical detectors e g the photomultiplier high voltages the amplification mode lin 3 or 4 decade logarithmic lower and upper le...

Страница 13: ...meter label To change parameter names click into the fields and enter the names by using the keyboard For each optical channel you may analyse three different pulse properties 1 The pulse height 2 the...

Страница 14: ...l other particles will not be recognized by the instrument The trigger parameter can be used for an efficient exclusion of unwanted particles from the analysis Example Assume vertebrate leukocytes in...

Страница 15: ...the instrument settings box see Fig 4 By using the Right Left buttons gain values are increased decreased In case linear amplification is used peaks are expanded to the right compressed to the left w...

Страница 16: ...decreased by clicking into the corresponding field in the instrument settings box and using the Right Left buttons see Fig 4 The count rate increases with elevated speed values If the speed is increas...

Страница 17: ...of small and unwanted background or noise signals below a threshold The L L value of a parameter can be increased decreased by clicking into the corresponding L L value field in the instrument setting...

Страница 18: ...educe smoke dust vibrations direct sunlight and direct neighbourhood of heatings as possible The installation room must be well ventilated and dry Temperature 15 30 C Humidity 20 85 relative non conde...

Страница 19: ...space shown from the rearside CyFlow rearside Screen Computer rearside Printer CCD Camera Interface Monitor Printer CCD Computer AC AC AC Air Sheath Mouse Keyboard Sheath Waste AC Waste CyFlow rearsid...

Страница 20: ...e supply sheath pressure and sample flow rate The Computer connections of the CyFlow space are located at the rear side of the instrument Connect the CyFlow space to the computer with the interface ca...

Страница 21: ...during sample run remove your sample and activate CLEAN several times In this way the flow cuvette is back flushed with sheath fluid Restart the sample and observe peak resolution again If the results...

Страница 22: ...lled water and a clean brush and flush with clean destilled water several times Remember cleanliness of sheath fluid reservoir is critical for proper operation If the CyFlow space will not be used for...

Страница 23: ...1 2001 Warning Laser light can be emitted if the protection cover for the laser beam is removed and the beam shutter is opened Therefore the system is marked with the following laser safety labels War...

Страница 24: ...entation control Particle Concentration Analysis True Volumetric Absolute Counting Particle Size and Fluorescence Distribution Analysis True Volumetric Based on precise counting and mechanical fluid v...

Страница 25: ...d precision syringe pump for contamination free sample transport Built in air pressure for sheath fluid Sheath fluid pressure is adjustable from 0 300 mbar Computer controlled Default setting 200 mBar...

Страница 26: ...d to 32 gates in free logical combinations Regions quadrants 1P ranges Gating Crosstalk On or offline gating and crosstalk compensation provide adjustments without need Compensation to rerun samples D...

Отзывы: