32
GB
Clearing calibrated data from the
product
Clean the electrode probe 1 and the test liquid
absorption rod 1a (see “Cleaning the electrode probe
and the test liquid absorption rod”).
Switch the product off and then on again (see “Power
on/off”).
pH
7 is shown in the display.
Press
/MODE
5 twice to enter the calibration
mode.
CAL
is flashing in the display 6 .
Press
CAL/HOLD
4 repeatedly until
CLr
is flashing
in the display 6 . Press
/MODE
5 .
NOTE:
If the button is not pressed within 2 minutes,
Err
starts flashing in the display 6 . Repeat all
previous steps of this section.