Operating the Generator
EN
15
4.4.3 Customer Settings
The customer settings menu contains all of the generator parameters that may be altered by the end user. The following example
demonstrates the method of altering a parameter, however it is recommended that none of the parameters are altered until their
functionality is fully understood.
Use the
and
keys to scroll through to the desired menu and press
.
Referring to the menu map select the desired.
The flashing cursor should be positioned over the “=” sign to indicate that the parameter may be
changed.
Use the
/
keys to change the parameter.
Press
to accept the changes or press
and
simultaneously to cancel the changes.
Press
and
simultaneously to return to the customer settings menu and then again to return to the main operating menu.
All settings marked in bold text are the default settings.
When enabled the O
2
alarm is over-ridden.
0 = Over-ride disabled
, 1 = Override Enabled
[OVR]
O
2
cell calibration menu.
Refer to section 4.7 for details on calibration.
Sets the purity level at which an oxygen fault is initiated.
Default Settings:
% Generators - 0.05% above the selected production purity
ppm Generators - 5ppm above the selected production purity
If the purity level exceeds the oxygen alarm level for a period longer than the alarm delay, the
oxygen alarm will be activated and the gas will be vented to atmosphere..
Delay Range = 0 – 600 Seconds,
Default = 60 Seconds
When enabled the bed cleaning cycles will run whenever the generator is powered up, comes out
of standby mode and economy mode.
0 = Disabled,
1 = Enabled
Enables the economy mode.
0 = Disabled,
1 = Enabled
When enabled the actuation of the Stop control will generate an alarm.
0 = Disabled
, 1 = Enabled
Sets the Full Scale Deflection value for the 4 – 20mA Analogue Output.
Sets the ppm O
2
cell calibrated offset value marked on the cell.
Note:
This value must be entered only when the cell is changed.
Sets the address for the generator when communicating on a network via the
RS485 MODBUS port.
Address range is
1
– 32
Sets the mode of control for the generator
1 = Local Start / Stop control
, 2 = Remote Start / Stop control via the digital input
3.7 Fault Relay
On Stop = 1