FRont PanEL DISPLayS
g
37
Parker Hannifin icountPD
Front panel displays
LED display parameters (ISo4406 / naS1638)
Start up
1. Once the icountPD has been connected to a regulated power supply, all four LEDs will
roll from top to bottom for a minimum of five seconds as the icountPD performs a system
diagnostic check.
2. The icountPD then automatically starts monitoring using the factory default test parameters.
Only if the Power on Mode is set to Auto (product default).
LED indication
The icountPD uses LED 1, 2 and 3 for the indication of ISO 4406; LED 2 alone is used for the
NAS1638 code. Individual code lights trigger according to user settings. The order of triggering is:
Solid green
■
■
trips at all codes below the set point (limit) code.
Flashing green
■
■
LED trips at the set point code.
Solid red
■
■
LED trips at the set point plus 1 code.
Flashing red
■
■
LED trips at the set point plus 2 codes.
LED 4 is reserved for the icountPD Moisture Sensor. Contact Parker Hannifin for details of its
specification.
ISo ExaMPLE
Assuming the limits on the icountPD have been set to 18/16/12 and the actual particles detected by
the icountPD is giving a ISO result of 20/17/11.
LED 1
■
■
flashes red
, 2 codes higher than the set limit.
LED 2 appears
■
■
solid red
(1 code higher),
LED 3 appears
■
■
solid green
(1 code lower).
naS ExaMPLE:
Assuming that the limit on the icountPD has been set to 7 and the actual particles detected by the
icountPD is giving a NAS result of 9.
LED 2
■
■
flashes red
(2 codes higher) .
LED MoIStURE SEnSoR InDICatoR (LED 4):
Solid green
■
■
trips at %RH (percentage Relative Humidity) level or below the set point (limit).
Solid red
■
■
trips above the %RH level.
LED 1
LED 2
LED 3
LED 4