SoFtWaRE
g
36
Parker Hannifin icountPD
With the Power-on Mode set to ‘Auto’ icountPD starts testing automatically when
the power is connected using the last setup parameters. With the Power-on Mode
set to ‘Manual’ icountPD becomes idle and requires the user to manually start
testing
.
SRI
S
et
R
eporting
I
nterval
SRI mm:ss
(0 to 3600 seconds (i.e. 0–1 hour); note
that 0 = No reporting)
The Reporting Interval controls how often icountPD sends results over the RS232
SRS
S
et
R
eporting
S
tandard
SRS iso
SRS nas
SSH
S
et limit relay
S
witch
H
ysteresis
2
SSH on
SSH off
SSS
S
et limit relay
S
witch
S
tate
2
SSS on
SSS off
SSU
S
et STI
S
ensor
U
sed
SSU yes
SSU no
SVM
S
et the
V
oltage
M
aximum range
3
SVM #
(3 = 0–3Vdc output
5 = 0–5Vdc output)
SWC
S
et
W
arning limit relay for
C
ontamination
2, 5
SWC on
SWC off
SWM
S
et
W
arning limit relay for
M
oisture
1, 2, 5
SWM on
SWM off
1
Command requires a Moisture sensor to be fitted to the icountPD
2
Command requires a Limit Relay to be fitted to the icountPD
3
Command requires a 0–5Vdc option to be fitted to the icountPD
4
Command requires digital display option to be fitted to icountPD
5
If the Limit Relay has been turned OFF for both Contamination monitoring and
Moisture sensing, the Limit Relay will not operate, but the alarm status is not
affected.
If the Limit Relay has been turned ON for both Contamination monitoring and
Moisture sensing, the Limit Relay will operate when any alarm condition is reached.