background image

Parker Hannifin Corporation  

Industrial Gas Filtration and Generation Division  
Haverhill, MA 

•  1-800-343-4048

 

Parker Zero Air Generators 

Models HPZA-3500, HPZA-7000, HPZA-18000, HPZA-30000

 

 

Bulletin TI-HPZA-3500H

 

 

     

 

 

 

 

 

 

 

 

 

Fuse Replacement

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Re-connect the replacement catalyst module by reversing the removal procedure. Hand tighten 
nuts on to ferrules and finish tightening with a wrench 1-1/2 turns for a tight seal. Prior to turning 
the zero air generator on, check for leaks by opening the compressed air supply to the unit. Use 
a soap or leak detecting solution to check the integrity of the piping and fittings.

 

Replace cover, without capturing any electrical wiring, and commence operation as detailed in 
the Start-Up section of this manual. The 

“Check System” indicator light on the front of the 

generator will flash for 2 to 3 minutes after start-up, until the circuit board and the catalyst 
module have been initialized. After the initialization period, the generator will commence 
operation from the warm-up stage. (

Note: 

if the 

“Check System” indicator light continues to blink 

after 5 to 10 minutes, please contact the factory.)

 

 
 

Occasionally, one or both of the fuses in the Zero Air Generator may burn out. The fuses are located 
in the power receptacle on the left-hand side of the generator.  

Before servicing the fuses, turn the generator off and disconnect the power cord from both 
the power supply and the generator power receptacle.

 

To access the fuses, use a small screwdriver to remove the fuse holder located in the power 
receptacle of the generator. In the Zero Air Generator, both the phase and neutral are fused 
separately. As a result of this configuration, both fuses should be checked any time fuse 
replacement is warranted. Replace either one or both fuses as necessary and re-assemble. 

To 

maintain the safety and performance integrity of the product, use only fuses of the size and 
type detailed in the specifications section of this bulletin.

 

Ordering Information 

Zero Air Generator

 

HPZA-3500

 

HPZA-7000

 

HPZA-18000

 

HPZA-30000

 

Replacement Prefilter, first stage

 

100-12-DX

 

100-12-DX

 

100-12-DX

 

100-12-DX

 

Replacement Prefilter, second stage

 

100-12-BX

 

100-12-BX

 

100-12-BX

 

100-12-BX

 

Replacement Final Filter

 

GS-050-05-95

 

GS-050-05-95

 

GS-050-05-95

 

GS-050-05-95

 

Replacement Catalyst Module

 

76810

 

76820

 

76811

 

76821

 

Maintenance Kit 

(1 yr)

 

MK7840

 

MK7840

 

MK7840

 

MK7840

 

Fuse 

(120 VAC)

 

13192

 

13192

 

13215

 

13215

 

Fuse 

(220 VAC)

 

13191

 

13191

 

13192

 

13192

 

 

 

 

Note: To ensure consistent product performance and reliability, use only genuine Parker replacement parts and filter cartridges.

 

 

Содержание HPZA-3500

Страница 1: ...e 1 are completely engineered systems which replace the use of inconvenient high pressure gas cylinders as a source of hydrocarbon free air The HPZA 30000 generators purify on site compressed air to a...

Страница 2: ...rs are equipped with float drains which automatically open to empty any liquids from inside the filter housing The drains are connected to 1 4 O D plastic tubing which discharges to atmosphere onthe r...

Страница 3: ...re taken into consideration Approximately 6 of space should be allowed along the top and sides of the generator to ensure proper ventilation Compressed Air The Parker Zero Air Generator requires a sou...

Страница 4: ...ion only Piping Components The inlet and outlet ports for the Parker Zero Air Generator are 1 4 NPT female and located on right side of the unit Inlet tubing and fittings should be clean and rated for...

Страница 5: ...stream from the generator Exposure of the generator to halogenated hydrocarbons will damage the catalyst module and void the warranty For additional information request the Analytical Gas Supplies cat...

Страница 6: ...carbon content of less than 0 05 ppm Models HPZA 3500 HPZA 7000 and HPZA 18000 a flow control device must be installed downstream from the generator to prevent downstream demand in excess of the gener...

Страница 7: ...Replace the GS membrane cartridge GS 050 05 95 and re assemble The time required to service all three stages of filtration integral to the Parker Zero Air Generator is approximately 15 minutes To prev...

Страница 8: ...tor may burn out The fuses are located in the power receptacle on the left hand side of the generator Before servicing the fuses turn the generator off and disconnect the power cord from both the powe...

Страница 9: ...generator is exceeded Overflow LED and yellow THC LEDs will illuminate An excessive pressure drop greater than 4 psig 0 3 bar may accompany an overflow situation If demand exceeds capacity install a f...

Страница 10: ...arg 40 psig 125 psig 2 8 barg 8 6 barg 40 psig 125 psig 2 8 barg 8 6 barg Min Max ambient temperature 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C Max relative h...

Страница 11: ...calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or i...

Страница 12: ...800 343 4048 Symbol Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 242 Neck Road Haverhill MA 01835 www parker com labgas Description Caution Refer to accompanying docum...

Отзывы: