background image

Parker Hannifin Corporation  

Industrial Gas Filtration and Generation Division  
Haverhill, MA 

•  1-800-343-4048

 

Parker Zero Air Generators 

Models HPZA-3500, HPZA-7000, HPZA-18000, HPZA-30000

 

 

Bulletin TI-HPZA-3500H

 

 

     

 

Recommended 
Accessories 

 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 

 
 

Pressure Regulator - 

To control the supply air pressure to the Zero Air Generator, a pressure 

regulator is needed. The 

72-130-V883 Pressure Regulator 

is a pressure regulator with 

½” NPT inlet 

and outlet ports, assembled with a pressure gauge.  For more information, please contact your local 
representative and request the Gas Management Supplies Catalog.

 

Flow Controller - 

If the output capacity of the Zero Air Generator is exceeded, the hydrocarbon 

content of the emergent gas may not meet specified levels.  To ensure zero air purity, install a flow 
controller designed for high purity applications downstream from the generator (if one is not 
integral to the GC design). The 

Parker Balston W-FM Flow Controllers 

provide accurate flow

 

rate control of the zero air stream. To determine which flow controller best meets the needs of the 
application, please contact your local representative and request the Gas Management Supplies 
Catalog.

 

76080 Halogenated Hydrocarbon Scrubber - 

If the compressed air supply to the generator is  subject 

to contamination from halogenated hydrocarbons, install a halogenated hydrocarbon scrubber directly 
upstream from the generator. Exposure of the generator to halogenated hydrocarbons will damage the 
catalyst module and void the warranty. For additional information, request the Analytical Gas Supplies 
catalog from your local representative.

 

 

Operation 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

Figure 5 - Front Panel Display

 

 

Start-up

 

Open the inlet compressed air supply valve (customer supplied, see Installation section) and adjust 
the inlet pressure to be between 40 psig and 125 psig (2.8 barg and 8.6 barg) using a (customer 
installed) pressure regulator. Adjust the outlet flow using a (customer installed) flow controller. Set the 
flow to the required flow rate for the instrument or application (see Specifications section at the end of 
this bulletin for generator flow capacities). 

Check delivery system for leaks. 

Finally, turn the 

generator on using the power switch on the right side of the unit.

 

During the 45-minute warm-up, the yellow 

“Start-Up” LED will remain illuminated, and the yellow 

“THC” LEDs will illuminate sequentially until the generator reaches the specified hydrocarbon removal 
level. After the warm-up period, the green 

“THC” LEDs will illuminate, showing a hydrocar bon level of 

<0.1 ppm. At this point, the Parker Zero Air Generator is ready to be brought online and supply zero-
grade air to downstream analytical equipment.

 

Note: 

If the 

“Start Up” LED goes dark, and the green “THC” LEDs do not illuminate, the yellow 

“Overflow” LED will illuminate. This indicates a downstream zero air demand which exceeds the 
capacity of the generator. Please see the Troubleshooting section at the end of this bulletin.

 

 

 

Содержание HPZA-3500

Страница 1: ...e 1 are completely engineered systems which replace the use of inconvenient high pressure gas cylinders as a source of hydrocarbon free air The HPZA 30000 generators purify on site compressed air to a...

Страница 2: ...rs are equipped with float drains which automatically open to empty any liquids from inside the filter housing The drains are connected to 1 4 O D plastic tubing which discharges to atmosphere onthe r...

Страница 3: ...re taken into consideration Approximately 6 of space should be allowed along the top and sides of the generator to ensure proper ventilation Compressed Air The Parker Zero Air Generator requires a sou...

Страница 4: ...ion only Piping Components The inlet and outlet ports for the Parker Zero Air Generator are 1 4 NPT female and located on right side of the unit Inlet tubing and fittings should be clean and rated for...

Страница 5: ...stream from the generator Exposure of the generator to halogenated hydrocarbons will damage the catalyst module and void the warranty For additional information request the Analytical Gas Supplies cat...

Страница 6: ...carbon content of less than 0 05 ppm Models HPZA 3500 HPZA 7000 and HPZA 18000 a flow control device must be installed downstream from the generator to prevent downstream demand in excess of the gener...

Страница 7: ...Replace the GS membrane cartridge GS 050 05 95 and re assemble The time required to service all three stages of filtration integral to the Parker Zero Air Generator is approximately 15 minutes To prev...

Страница 8: ...tor may burn out The fuses are located in the power receptacle on the left hand side of the generator Before servicing the fuses turn the generator off and disconnect the power cord from both the powe...

Страница 9: ...generator is exceeded Overflow LED and yellow THC LEDs will illuminate An excessive pressure drop greater than 4 psig 0 3 bar may accompany an overflow situation If demand exceeds capacity install a f...

Страница 10: ...arg 40 psig 125 psig 2 8 barg 8 6 barg 40 psig 125 psig 2 8 barg 8 6 barg Min Max ambient temperature 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C 60 F 16 C 100 F 38 C Max relative h...

Страница 11: ...calling for assistance WARRANTY NORTH AMERICA ONLY FOR INFORMATION CONTACT YOUR LOCAL REPRESENTATIVE Parker Hannifin guarantees to the original purchaser of this product that if the product fails or i...

Страница 12: ...800 343 4048 Symbol Parker Hannifin Corporation Industrial Gas Filtration and Generation Division 242 Neck Road Haverhill MA 01835 www parker com labgas Description Caution Refer to accompanying docum...

Отзывы: