INSTALLATION, OPERATION, AND MAINTENANCE MANUAL ALIGN™ MULTI-GAS GENERATOR
8
1
For purity at specific flow rates and pressures please contact Parker Hannifin.
2
For humidity and temperature information outside of this range please contact Parker Hannifin.
3
For altitudes outside of this range please contact Parker Hannifin.
3.3 Overview of Components
System Specifications Cont
Outlet Gas Specifications Cont’
Phthalates
None
Max Tank Operating Pressure
115±2 psig (7.9±0.1 barg, 793 kPa)
Gas Outlets
1/4’’ NPT
Environmental Requirements
Min/Max Ambient Temperature
60°F/ 85°F (16°C/29°C)
Max Ambient Relative Humidity
2
43% at 86°F
Max Recommended Operating Altitude
3
2000 m
Electrical Requirements
Electrical Supply for ALIGN-MG-NA, ALIGN-MG-JA
230VAC ± 15%, 50/60 Hz, 1 Phase, 20A, NEMA 6-20R receptacle
Electrical Supply for ALIGN-MG-WD
230VAC ± 15%, 50/60 Hz, 1 Phase, 32A, IP44 receptacle
Power Consumption
15A@230VAC, 2500 Watts
Fuses (F1, F2)
JJS Class T, 600V, 25A, 14.2 mm Dia x 38.1 mm
IOM_PKR_Align Multi-Gas Generator_022021.indd 8
IOM_PKR_Align Multi-Gas Generator_022021.indd 8
2/17/2021 1:11:43 PM
2/17/2021 1:11:43 PM