INSTALLATION, OPERATION, AND MAINTENANCE MANUAL ALIGN™ MULTI-GAS GENERATOR
Replace front and rear panels before proceeding.
4.4. Choosing an Installation Location
This equipment is not suitable for use in any hazardous, flammable, or explosive environments.
Keep the generator away from excessive heat and open flame.
The accumulation of nitrogen can displace oxygen thereby creating an asphyxiation hazard.
Always ensure that the equipment is operated in a well-ventilated area.
Locate the generator as close to the instrument as possible following the requirements given below:
4.4.1. Environmental & Space Requirements
The generator is for indoor use only. Do not install outdoors. The generator must be installed
in a climate-controlled room, meeting the environmental specifications set in Section 3.2.
It is critical that nothing is placed within 12 inches of the exhaust vent, located at the left rear
side of the generator. An optional Vent Kit is available to route exhaust heat away from the
generator, see Section 8 for purchasing information.
Do not install the generator where hot exhaust air from other equipment may be drawn into
the intake at the rear of the generator. The area should be free of excessive dust, toxic or
flammable gases, solvent fumes, and moisture.
Figure 7
Figure 8
Figure 9
Figure 10
IOM_PKR_Align Multi-Gas Generator_022021.indd 11
IOM_PKR_Align Multi-Gas Generator_022021.indd 11
2/17/2021 1:11:57 PM
2/17/2021 1:11:57 PM