Bulletin TI-NITROFLOW TG2-G
Installation, Operation and Maintenance Manual
TI-NITROFLOWTG2
www.labgasgenerators.com
16
1-800-343-4048
Carbon Modules
Particulate Filters
During the 10,000 hour service interval only the inlet Carbon
Module (B06-0264) closest the rear of the generator is replaced
(see figure 10). All three carbon modules are replaced during the
30,000 hour service. Please note that the carbon modules will only
mount to the brackets in one orientation.
1
Using a 1” wrench, remove the tie nut at the base of the bowl (see Figure 40).
2
Twist the bowl loose from the head of the assembly and remove carefully as not to damage or
lose the o-ring (see Figure 41).
3
Pull off the element retainer at the base of the filter, and replace the spent filter element with a
new one (see Figure 42).
4
Reverse the steps to re-assemble. Make sure the o-ring is sealed in the groove before
reinstalling the bowl. Hand tighten, plus 1/4 turn using the 1” wrench.
Figure 39
Figure 40
Figure 41
Figure 42
1
Remove the top panel from the NitroFlow TG2 Generator Series by removing two screws with
a phillips head screwdriver and pulling top panel to the rear.
2
Disconnect the plastic tubing from the press fittings on the inlet and outlet ports of the module
3
Release the hook and loop strap that clamps the module to the unit (see Figure 39). Slide the
module out of the bracket.
4
Replace with a new carbon module, noting the correct orientation.
5
Slide the new module into the bracket fully and re-attach the hook and loop strap.
6
Re-insert inlet and outlet tubing. Use a soap or a leak detecting solution to check for leaks
when pressurizing (re-starting) the unit.
Generator must be set to “OFF” and pressure gage must read zero before filter replacement.