Bulletin TI-H2PEMPD-E
Installation, Operation and Maintenance Manual
Hydrogen Gas Generator H2 PEMPD series
2
1-800-343-4048
www.labgasgenerators.com
1.1 Markings and Symbols
The following markings and international symbols are used on the equipment or within this manual:
Table 1
2 Description
This generator will produce a constant stream of high purity hydrogen at a predetermined flow rate and pressure when connected to
a suitable power supply and fed with a suitable quality of deionized water. It is suitable for use in laboratories and light industrial
environments and is nonhazardous for transportation purposes.
In order to guarantee the optimum efficiency of the PEM cell, this generator must be installed and running within three months of
dispatch from Parker Hannifin. Failure to do this may invalidate the warranty. The generator will perform a 60 minute (1 hour)
initialization sequence when powered for the first time. This sequence, which cannot be aborted, is necessary to guarantee the
correct hydration of the cell.