Bulletin TI-H2PEMPD-E
Installation, Operation and Maintenance Manual
Hydrogen Gas Generator H2 PEMPD series
22
1-800-343-4048
www.labgasgenerators.com
6 Error Messages
When an error occurs the LCD will cycle between the default menu and the error message. In addition to the error messages the
generator will provide a visual and audible indication using the System Check LED, Water Bottle Indicator and its integral sounder.
LCD Display
Cause
Generator Status
Action
Water below the
midpoint
On
Fill with Deionized,
ASTM II, >1MΩ, <1µS,
filtered to <100µm
Water below the
minimum point
Off
Fill with Deionized,
ASTM II, >1MΩ, <1µS,
filtered to <100µm
Water conductivity HIGH
On
Drain and refill with
Deionized, ASTM II,
>1MΩ, <1µS, filtered to
<100µm
Water conductivity HIGH
Off
Drain and refill with
Deionized, ASTM II,
>1MΩ, <1µS, filtered to
<100µm
Cell voltage HIGH
Off
Reset or hard reset.
Contact tech support for
cell or PSU replacement
if the problem persists.
Cell voltage LOW
Off
Reset or hard reset.
Contact tech support for
cell or PSU replacement
if the problem persists.
Cell current HIGH
Off
Reset or hard reset.
Contact tech support for
cell or PSU replacement
if the problem persists.
Cell current LOW
Off
Reset or hard reset.
Contact tech support for
cell or PSU replacement
if the problem persists.
Faulty H2 line internal
connection
Off
Check all H2 line
connections
Cell low pressure
Off
Check/eliminate overflow
condition
Generator/Cell operating
at >100% capacity
Off
Check/eliminate overflow
condition
FILL WATER
ACT 100%
WATER EMPTY
ACT 100%
CHANGE WATER
ACT 100%
CONDUCTIVITY
ACT 100%
HIGH VOLTAGE
ACT 100%
LOW VOLTAGE
ACT 100%
HIGH CURRENT
ACT 100%
LOW CURRENT
ACT 100%
INTERNAL LEAK
ACT 100%
LOW PRESSURE
ACT 100%
OVER RUN
ACT 100%