has been saturated with an EPA-registered 

low-level (HIV/HBV claim) to intermediate-

level (tuberculocidal claim) hospital 

disinfectant, wiped of chemical residue with 

a water-damp paper towel, and dried. 

Utilize the surface disinfecting protocol of 

the disinfectant manufacturer.

• Since sterilization does not remove 

accumulated debris, Scaling Tips and 

Wrench must be washed in soap and 

water either manually or in an ultrasonic 

cleaner, rinsed in running water and dried 

prior to bagging. Failure to adequately 

remove debris, especially from the narrow 

water passages in the tip, will result in 

inadequate sterilization.

• Scaling Tips and Wrench may be 

sterilized in any conventional steam 

autoclave following manufacturer’s 

instructions. A typical gravity steam cycle 

is 270° F (132º C) for 15 minutes at a 

minimum of 15 psi, followed by a 

15-minute minimum cool-down period. 

After sterilization, inspect the device in 

the autoclave bag for integrity. If 

suspicion about the item exists, discard it 

and order a replacement from Parkell or 

your dealer.

TROUBLESHOOTING

  Unit does not operate (power “ON” 

indicator does not light):

• Transformer cable disconnected from wall 

outlet or scaler. Reconnect properly.

• Wall outlet not live. Contact electrician 

for repair.

• Unit fuse has failed. Contact Parkell.

  Tip does not vibrate properly: 

• Faulty, damaged or worn tip. Replace.
• Tip not correctly threaded onto handpiece. 

Re-attach properly.

• Power control not correctly adjusted. 

Re-adjust until correct.

• Use of non-compatible, non-Parkell tip.

  Excess heat at handle or tip: 

• Faulty, damaged or worn tip. Replace.
• Too little water flow. Tip water passage 

may be blocked. Inspect and clear 

blockage, or replace tip.

• Power too high for the amount of water. 

Increase water flow.

• Excessive hand pressure applied. 

Correct technique.

• Heated water being used. Use room 

temperature water only.

   No water spray when foot-pedal is 

depressed:

• Water line blocked or kinked. Correct.
• Water passage in the scaling tip clogged. 

Inspect and clear blockage, or replace tip.

• Water filter clogged. Replace.
• No water supply connected. Correct.

  Tip does not go on handpiece:

• Threads damaged. Replace tip or 

handpiece.

USER MAINTENANCE AND 

AUTHORIZED SERVICE

• The Parkell TurboPIEZO requires very little 

maintenance for years of trouble-free 

operation. There are no user-serviceable 

parts inside the control unit or the foot 

pedal. Opening the cases will void all 

warranties. Authorized Parkell personnel 

must make all internal repairs.

• Parkell urges all scaler users to monitor the 

condition of the water filter periodically and 

change the filter whenever noticeable 

darkening of the disk occurs, or when water 

flow decreases. The most common reason 

for service for ultrasonic scalers is failure 

by the user to periodically change the water 

filter. This will result in contamination and 

clogging of the internal water components, 

as well as costly repair service.

• The ultrasonic tips, power supply, water 

filter and water hose are all detachable and 

may be replaced by the user. Contact 

Parkell for a list of authorized tips and 

available replacement parts.

• Within the US, all repairs must be made by 

Parkell. We maintain a complete service 

and parts facility in our factory at 300 

Executive Drive, Edgewood, NY 11717. 

After calling our factory for advice, 

equipment needing service in the US 

should be returned, freight pre-paid, via 

approved common carrier (e.g. USPS, 

UPS, FedEx), and adequately insured. 

Return all accessories with the unit, and 

include an explanation of the problem. Pack 

in the original box, add plenty of cushioning 

material, and overbox the unit during 

Содержание TurboPIEZO

Страница 1: ...lectrical devices the unit should not be immersed in water or other liquids Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device after it has...

Страница 2: ...Straight hD693 Perio hD695 Burnett Power Tip 1 Scaler Tip Wrench hD691 2 Inline water filters one attached to water line hD419 bag of replacement filters 1 Transformer with cables hD408 120V hD403 23...

Страница 3: ...e OUT position CONTROLLING THE POWER OUTPUT The TurboPIEZO gives you several ways to adjust the scaling power 1 Use the Power Control Knob on the main power unit to set the initial scaling power for t...

Страница 4: ...sink or cuspidor until desired mist is obtained For maximum patient comfort use a good flow of water as a coolant lubricant and to flush out debris Hold the handpiece in a comfortable pen grasp To re...

Страница 5: ...the amount of water Increase water flow Excessive hand pressure applied Correct technique Heated water being used Use room temperature water only No water spray when foot pedal is depressed Water lin...

Страница 6: ...1 249 1134 or email our Technical Support Service at techsupport parkell com WARRANTY For full Warranty and Terms of Use information please see www parkell com Parkell s Quality System is certified to...

Страница 7: ......

Страница 8: ...dental device was designed manufactured and is serviced in the United States of America by 300 Executive Drive Edgewood NY 11717 USA Toll Free 800 243 7446 Phone 631 249 1134 E mail info parkell com W...

Отзывы: