parkell TurboPIEZO Скачать руководство пользователя страница 3

2. Plug the AC transformer’s three-pronged 

power cord into a grounded electrical 

outlet.

 

SCALING TIPS FOR THE  

TURBOPIEZO SCALER

The TurboPIEZO scaler comes with 3 tips 

included with the basic scaler unit. 

Replacements are available from Parkell. 

Prior to use, the tips should be sterilized as 

described later. Tips are easily threaded on 

and off the handpiece using the Scaler Tip 

Wrench. Tips should be tightened clockwise 

firmly with the wrench until it disengages. 

Tips are loosened by turning the wrench 

counterclockwise. Bent, worn or blunt tips 

will scale poorly and generate excess heat, 

and should be replaced with new.

IMPORTANT INFORMATION 

REGARDING USE OF  

NON-PARKELL TIPS:

Parkell piezoelectric scaler tips may be 

compatible with other “EMS-compatible” 

brands of tips. However, for optimal 

performance of your scaler, we recommend 

using Parkell ultrasonic tips, which have been 

tested to function in our unit. Parkell cannot 

guarantee that all non-Parkell tips available 

on the market today or in the future will work 

properly with our scalers. For more 

information, contact Parkell.

TURNING THE SCALER ON AND OFF

Turn the unit on by pressing the “ON/OFF” 

button on the front panel to the “IN” position. 

The Green “ON” LED will illuminate. The 

scaler may now be activated by pressing 

down on the foot-pedal. When the foot-pedal 

is pressed, the Green “ON” LED will get 

brighter, indicating that the unit is active. 

When the foot-pedal is released, the Green 

“ON” light will return to original brightness. To 

turn the unit off, press the “ON/OFF”button to 

the “OUT” position.

CONTROLLING THE POWER OUTPUT

The TurboPIEZO gives you several ways to 

adjust the scaling power.
1. Use the Power Control Knob on the main 

power unit to set the initial scaling power 

for the procedure at hand. Turn the knob in 

a clockwise direction to increase the power, 

or counterclockwise to decrease the power.

• Usually, the Power Control Knob is set at 

the “Medium” setting to start. When the 

foot-pedal is slightly depressed, the tip will 

begin vibrating at Medium power. 

• The “PERIO” button reduces the range of 

the power control knob. When this switch 

is pressed, it lights up. Now, relatively large 

movements of the Power Control Knob will 

produce small changes in tip power. This 

permits very subtle adjustments that are 

useful for subgingival scaling.

2. Use the “Turbo Mode” for short-term 

increases in power during scaling. Press 

the foot-pedal using increased foot 

pressure and the yellow “TURBO” light 

will illuminate. This will instantly boost 

scaling power to a point midway between 

the current setting and the scaler’s 

maximum power.

• Engaging the Turbo feature at a low power 

setting causes a significant boost in power. 

• Engaging the Turbo feature at high power 

will produce very little change.

• Since water flow is not affected by “Turbo 

Mode”, extended procedures may require a 

water adjustment.

• Partially lifting your foot off the foot-pedal 

will return the scaler to normal, “non-

Turbo” mode, and the yellow “TURBO” 

light will go out.

CONTROLLING THE WATER FLOW

Observing the arrow printed on the control 

panel, turn the water control counter-

clockwise to increase the flow, or clockwise to 

decrease water flow. If water does not flow 

through the tip when the foot-pedal is 

depressed, consult the Troubleshooting 

Section of this manual.

Содержание TurboPIEZO

Страница 1: ...lectrical devices the unit should not be immersed in water or other liquids Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device after it has...

Страница 2: ...Straight hD693 Perio hD695 Burnett Power Tip 1 Scaler Tip Wrench hD691 2 Inline water filters one attached to water line hD419 bag of replacement filters 1 Transformer with cables hD408 120V hD403 23...

Страница 3: ...e OUT position CONTROLLING THE POWER OUTPUT The TurboPIEZO gives you several ways to adjust the scaling power 1 Use the Power Control Knob on the main power unit to set the initial scaling power for t...

Страница 4: ...sink or cuspidor until desired mist is obtained For maximum patient comfort use a good flow of water as a coolant lubricant and to flush out debris Hold the handpiece in a comfortable pen grasp To re...

Страница 5: ...the amount of water Increase water flow Excessive hand pressure applied Correct technique Heated water being used Use room temperature water only No water spray when foot pedal is depressed Water lin...

Страница 6: ...1 249 1134 or email our Technical Support Service at techsupport parkell com WARRANTY For full Warranty and Terms of Use information please see www parkell com Parkell s Quality System is certified to...

Страница 7: ......

Страница 8: ...dental device was designed manufactured and is serviced in the United States of America by 300 Executive Drive Edgewood NY 11717 USA Toll Free 800 243 7446 Phone 631 249 1134 E mail info parkell com W...