CLINICAL PRECAUTIONS

• Protect patient’s eyes, lips, cheeks, tongue 

or other vulnerable tissues from contact with 

the vibrating scaler tip.

• Protect the patient’s clothing from water 

damage when using this device.

• As with all dental procedures, clinicians 

should use universal precautions (i.e., wear 

face mask, eyewear or face shield, gloves 

and protective gown). 

• Keep the long axis of the tip parallel to the 

long axis of the tooth while maintaining 

gentle contact with the root to “wipe” 

deposits from the tooth. Prolonged 

perpendicular contact between the tip and 

the tooth will damage the tooth by gouging it.

CONFORMANCE TO STANDARDS

Parkell’s TurboPIEZO is TUV listed and 

conforms to IEC 60601-1, 60601-1-2 and 

CAN/CSA C22.2 No. 601.1. Parkell’s quality 

system is certified to ISO13485.

SPECIFICATIONS

• 

Power:

 120VAC model – 50/60 Hz, 35VA 

 

230VAC model – 50 Hz, 35VA

• 

Operating temperature:

  

Between 50°F and 100°F (10°C - 38°C)

• 

Weight:

 30 oz

• 

Size:

 1 ⅝” H x 5 ½” W x 7 ½” D

• 

Handpiece/Tip operating frequency:

 29 KHz

ADDITIONAL SPECIFICATIONS

• Protection against electric shock: Class 1, 

Type B applied part.

• Protection against ingress of liquids:  

Foot Switch & Scaler: IPX1 (drip proof); 

Power Supply: IPX0 (Ordinary).

• Mode of operation of equipment: Intermittent 

– 10 minutes on, 5 minutes off.

• Operating conditions: 60-86°F (15-30°C), 

10-80% RH (non-condensing).

• Transport and storage conditions: 50-100°F 

(10-40°C), 10-80% RH (non-condensing). 

TURBOPIEZO IS SUPPLIED WITH:

(1)   Piezoelectric Ultrasonic Scaler Unit with 

detachable handpiece, attached foot 

control and water line (with water filter and 

male quick-connect coupler attached); 

120VAC = 

h

D690-110 

230VAC = 

h

D690-230

(3)  Tips:  

h

D692-Universal Straight 

h

D693-Perio 

h

D695-Burnett Power-Tip

(1)   Scaler Tip Wrench (

h

D691). 

(2)   Inline water filters (one attached to water 

line) (

h

D419 - bag of replacement filters). 

(1)   Transformer with cables (

h

D408 (120V); 

h

D403 (230V). 

(1)   Instructions for Use.

INSTALLING YOUR SCALER

Locate the device where the control panel 

will be easy to reach during scaling 

procedures, and where the water filter at the 

rear of the unit may be periodically changed 

without difficulty. The scaler requires access 

to a grounded electrical outlet and a source 

of drinking-quality water. The device and its 

separate transformer generate a minimal 

amount of heat. Avoid covering them, to allow 

normal cooling.

WATER CONNECTION

Before plugging in the electrical cord, 

connect the scaler’s water line to a drinking-

quality water supply (15-35 psi), clean and 

free of sediment. An external, office-wide 

water filter is recommended to minimize 

frequent changes of the device’s in-line water 

filter. The male quick-connect that comes on 

the end of the water hose is the standard 

fitting presently used in the dental industry. If 

you are replacing a scaler that uses an 

Adec-type or other type of connection, 

remove the old fitting from the unit hose and 

attach it to the new water line. Check for 

leaks at all hose and filter connections before 

use, and tighten if necessary.

IMPORTANT PRECAUTIONS

The water supply line to the scaler should 

ALWAYS be turned off whenever the device is 

being connected, disconnected or is not in 

use. The foot-pedal should ALWAYS be placed 

where it will not accidentally be pressed when 

the device is not in use.

ELECTRICAL CONNECTION

The TurboPIEZO must only be powered via 

the Transformer Assembly that is supplied with 

the device by Parkell.
1. Plug the low voltage, round transformer 

power cord connector into the back of the 

scaler, with the arrow on top.

Содержание TurboPIEZO

Страница 1: ...lectrical devices the unit should not be immersed in water or other liquids Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device after it has...

Страница 2: ...Straight hD693 Perio hD695 Burnett Power Tip 1 Scaler Tip Wrench hD691 2 Inline water filters one attached to water line hD419 bag of replacement filters 1 Transformer with cables hD408 120V hD403 23...

Страница 3: ...e OUT position CONTROLLING THE POWER OUTPUT The TurboPIEZO gives you several ways to adjust the scaling power 1 Use the Power Control Knob on the main power unit to set the initial scaling power for t...

Страница 4: ...sink or cuspidor until desired mist is obtained For maximum patient comfort use a good flow of water as a coolant lubricant and to flush out debris Hold the handpiece in a comfortable pen grasp To re...

Страница 5: ...the amount of water Increase water flow Excessive hand pressure applied Correct technique Heated water being used Use room temperature water only No water spray when foot pedal is depressed Water lin...

Страница 6: ...1 249 1134 or email our Technical Support Service at techsupport parkell com WARRANTY For full Warranty and Terms of Use information please see www parkell com Parkell s Quality System is certified to...

Страница 7: ......

Страница 8: ...dental device was designed manufactured and is serviced in the United States of America by 300 Executive Drive Edgewood NY 11717 USA Toll Free 800 243 7446 Phone 631 249 1134 E mail info parkell com W...