CLINICAL PRECAUTIONS
• Protect patient’s eyes, lips, cheeks, tongue
or other vulnerable tissues from contact with
the vibrating scaler tip.
• Protect the patient’s clothing from water
damage when using this device.
• As with all dental procedures, clinicians
should use universal precautions (i.e., wear
face mask, eyewear or face shield, gloves
and protective gown).
• Keep the long axis of the tip parallel to the
long axis of the tooth while maintaining
gentle contact with the root to “wipe”
deposits from the tooth. Prolonged
perpendicular contact between the tip and
the tooth will damage the tooth by gouging it.
CONFORMANCE TO STANDARDS
Parkell’s TurboPIEZO is TUV listed and
conforms to IEC 60601-1, 60601-1-2 and
CAN/CSA C22.2 No. 601.1. Parkell’s quality
system is certified to ISO13485.
SPECIFICATIONS
•
Power:
120VAC model – 50/60 Hz, 35VA
230VAC model – 50 Hz, 35VA
•
Operating temperature:
Between 50°F and 100°F (10°C - 38°C)
•
Weight:
30 oz
•
Size:
1 ⅝” H x 5 ½” W x 7 ½” D
•
Handpiece/Tip operating frequency:
29 KHz
ADDITIONAL SPECIFICATIONS
• Protection against electric shock: Class 1,
Type B applied part.
• Protection against ingress of liquids:
Foot Switch & Scaler: IPX1 (drip proof);
Power Supply: IPX0 (Ordinary).
• Mode of operation of equipment: Intermittent
– 10 minutes on, 5 minutes off.
• Operating conditions: 60-86°F (15-30°C),
10-80% RH (non-condensing).
• Transport and storage conditions: 50-100°F
(10-40°C), 10-80% RH (non-condensing).
TURBOPIEZO IS SUPPLIED WITH:
(1) Piezoelectric Ultrasonic Scaler Unit with
detachable handpiece, attached foot
control and water line (with water filter and
male quick-connect coupler attached);
120VAC =
h
D690-110
230VAC =
h
D690-230
(3) Tips:
h
D692-Universal Straight
h
D693-Perio
h
D695-Burnett Power-Tip
(1) Scaler Tip Wrench (
h
D691).
(2) Inline water filters (one attached to water
line) (
h
D419 - bag of replacement filters).
(1) Transformer with cables (
h
D408 (120V);
h
D403 (230V).
(1) Instructions for Use.
INSTALLING YOUR SCALER
Locate the device where the control panel
will be easy to reach during scaling
procedures, and where the water filter at the
rear of the unit may be periodically changed
without difficulty. The scaler requires access
to a grounded electrical outlet and a source
of drinking-quality water. The device and its
separate transformer generate a minimal
amount of heat. Avoid covering them, to allow
normal cooling.
WATER CONNECTION
Before plugging in the electrical cord,
connect the scaler’s water line to a drinking-
quality water supply (15-35 psi), clean and
free of sediment. An external, office-wide
water filter is recommended to minimize
frequent changes of the device’s in-line water
filter. The male quick-connect that comes on
the end of the water hose is the standard
fitting presently used in the dental industry. If
you are replacing a scaler that uses an
Adec-type or other type of connection,
remove the old fitting from the unit hose and
attach it to the new water line. Check for
leaks at all hose and filter connections before
use, and tighten if necessary.
IMPORTANT PRECAUTIONS
The water supply line to the scaler should
ALWAYS be turned off whenever the device is
being connected, disconnected or is not in
use. The foot-pedal should ALWAYS be placed
where it will not accidentally be pressed when
the device is not in use.
ELECTRICAL CONNECTION
The TurboPIEZO must only be powered via
the Transformer Assembly that is supplied with
the device by Parkell.
1. Plug the low voltage, round transformer
power cord connector into the back of the
scaler, with the arrow on top.