FAULT
DESCRIPTION
THINGS TO CHECK
SOLUTION TO
PROBLEM
IF STILL NOT
WORKING
1. THE PUMP IS
SHOWING NO
INDICATORS TO SAY
THAT IT IS ON. . .
MAINS ARE PLUGGED IN
PLUG IN
RETURN PUMP
TO SERVICE
PROVIDER
2. THE LOW
PRESSURE LIGHT IS
CONSTANTLY
FLASHING AND THE
ALARM IS SOUNDING
3. THE PUMP IS
ON BUT NOT
INFLATING THE
SEAT
RETURN
SYSTEM TO
YOUR PROVIDER. . .
RETURN
SYSTEM TO
YOUR SERVICE
PROVIDER
MAINS ARE PLUGGED IN
TURN ON
FRONT SWITCH IS ON
TURN ON
FUSES ON THE SIDE
PUMP ARE NOT BLOWN
REPLACE WITH A 1 AMP
FUSE
FUSE IN THE PLUG
BLOWN
REPLACE WITH A 3 AMP
FUSE
THE HOSES ARE ALL
CONNECTED
RELOCATE HOSES
THE C.P.R. IS IN PLACE
RELOCATE C.P.R.
THERE ARE NO LEAKS IN
THE SEAT SYSTEM
RETURN MATTRESS TO
YOUR SERVICE PROVIDER
DISCONNECT HOSES
FROM PUMP TO SEE IF
AIR IS COMING OUT
RETURN PUMP TO YOUR
SERVICE PROVIDER
THE HOSES ARE ALL
CONNECTED
RECONNECT HOSES
THE C.P.R. IS IN PLACE
RELOCATE C.P.R.
THERE ARE NO OBVIOUS
AIR LEAKS IN THE SYSTEM
SEAT SYSTEM
RETURN YOUR SEAT
TO YOUR SERVICE
PROVIDER
CHECK THE PRESSURE
IS IN THE SUGGESTED
RANGE FOR THE WEIGHT
OF THE PATIENT
TURN THE PRESSURE UP
TO WITHIN THE CORRECT
WEIGHT RANGE
THERE ARE NO KINKS
IN THE TUBING RUNNING
DOWN THE SIDE OF
THE SEAT
UNTWIST ANY KINKS
FOUND
4. THE SYSTEM
DOES NOT
APPEAR TO BE
ALTERNATING
THERE ARE NO
KINKS IN THE
TUBING RUNNING
DOWN THE SIDE
UNTWIST ANY
KINKS FOUND
RETURN SYSTEM TO
YOUR SERVICE
PROVIDER
5. THE PUMP IS
OPERATING
NOISILY
THE PUMP IS NOT
RESTING AGAINST
A SOLID SURFACE
RETURN PUMP TO
YOUR SERVICE
PROVIDER
6. Troubleshooting
If your problem cannot be resolved using the above information. Please
cantact Park House Healthcare.
1. General Safety
Before you use your Eclipse Seating system, please familiarise yourself
with the information in this user manual. Keep this manual in a handy place
for future reference.
For you safety and the ef
fi
cient performance of the system the follwing
precautions should be observed:
-
MAXIMUM PATIENT WEIGHT 25 stones / 159 kgs
-
Do not expose the pump to liquids of any kind
-
Always protect the systems from open
fl
ames
-
Avoid the use of Phenol based solutions when cleaning
-
Store the system in a clean and dry environment
-
Do not store in direct sunlight
NOTE: The pump in this system is
fi
tted with a functional earth, and is
labelled accordingly.
There is a possibility of electro-magnetic interference with other
devices. If so increase distance between devices.
Double insulation
See instructions
for use
Type BF equipment
As labelled
accordingly
}