an ambient temperature of 23 °C and 50% relative humidity.
2 ml salbutamol (1 mg/1 ml) were nebulised. The Jet Flow was
measured to be 4.6 l/min.
Nozzle insert (orange)
Nominal com-
pressor flow
(4.6 l/min –
1.5 bar)
MMAD [µm]
8
3.0
GSD
9
2.18
Aerosol fraction [% > 5 µm]
66.6
Aerosol fraction [% < 2 µm]
28.7
Aerosol fraction [% > 2 µm < 5 µm]
38.0
Aerosol fraction [% > 5 µm]
33.4
Aerosol output [ml]
0.78
Aerosol output rate [ml/min]
0.15
Residual volume [ml] (gravimetric)
0.95
Percentage of fill volume emitted per minute [%/min] 7.54
7
MISCELLANEOUS
7.1 Disposal
Compressor
This product falls within the scope of the European Council Dir-
ective on Waste Electrical and Electronic Equipment (WEEE)
10
.
Accordingly, this product must not be disposed of with do-
mestic waste. The disposal regulations prevailing in the re-
8) MMAD = Mass Median Aerodynamic Diameter
9) GSD = Geometric Standard Deviation
10) Directive 2012/19/EU of the EUROPEAN PARLIAMENT AND THE
EUROPEAN COUNCIL of July 4, 2012 on waste electrical and electronic
equipment.