REPROCESSING
DANGER
Danger of electrocution
Liquids can conduct electricity, thereby posing a risk of elec-
tric shock.
• Always switch the compressor off and disconnect the
power plug from the mains socket before starting to clean
it.
If the nebuliser system is used
by only one patient
, follow the
instructions for cleaning and disinfecting in the section "Single
patient use" [see: Single patient use, page 25].
If the nebuliser system is used
by several patients
, follow the
instructions for cleaning and disinfecting in the section "Multiple
patient use" [see: Multiple patient use, page 29].
If the nebuliser system is used
in professional environments
,
follow the information on reprocessing included in the appendix
at the end of these instructions for use.
4.1 Single patient use
Cleaning and disinfection cycles
Compressor housing
Clean when visibly dirty
Nebuliser (without
PARI SINUS tubing
system)
– Cleaning immediately after every use.
– Disinfection at least once a week.
In the case of an acute infection or for
high risk groups (e.g., cystic fibrosis
patients) at least once a day.
Air filter
Replace after 200 operating hours
(approx. 1 year)
– 25 –