E. How To Clean and Maintain Your eRapid
9 Questions about eRapid? Call 1-866-335-6943
d.
Remove the Aerosol Head.
Be careful to touch
ONLY
the plastic outer ring of the Aerosol Head,
press the 2 flexible plastic arms on the side of the
Aerosol Head and remove
(Figure E2d). Do Not
touch the center metal part of the Aerosol Head.
Step E3:
Rinse each of the disassembled parts
under
warm (approximately 104°F/ 40°C) running tap
water to remove residue (for approximately 10
seconds).
Note
:
there is no need to
disassemble the Medication Reservoir from the
aerosol chamber for cleaning.
Step E4:
Rinse each side of the Aerosol Head for at least
10 full seconds (Figure E4).
Never use a brush or
any other object to clean the Aerosol Head.
Thorough rinsing of both sides of the Aerosol
Head helps with proper operation and prevents
clogging.
Step E5:
Soak all parts of the Handset in warm soapy
water for 5 minutes
made by adding about ¼
teaspoon of clear liquid dish soap and 1 quart of
warm tap water to a clean tub or bowl (
Figure E5
).
Step E6:
Inspect
all parts to confirm they are clean. If they
are not clean, soak them in warm soapy water for
an additional 5 minutes with gentle agitation.
Step E7:
Rinse all parts thoroughly
under warm running
tap water
(Figure E7).
Step E8:
Air dry parts
on a clean, lint-free cloth or paper
towel (
Figure E8
). After the parts are completely
dry, place them in a bag or cover them with a lint-
free cloth or paper towel for storage. Re-assemble
just before treatment.
Arms
E2d
E4
E7
E5
E8